Lutron Electronics SL-4010 Operation Manual Download Page 4

Time Weighting

Default to " Fast " time weighting

characteristics.

 * " Fast time weighting " is simulated the

human ear response character.

Accuracy

Characteristics of " A " frequency

(23± 5  )

weighting network meet IEC 61672 

class 2.

Under 94 dB input signal,  the accuracy

are :

31.5 Hz

± 3.5 dB

63

Hz

± 2.5 dB

125 Hz

± 2.0 dB

250 Hz

± 1.9 dB

500 Hz

± 1.9 dB

1 K Hz

± 1.4 dB

2 K Hz

± 2.6 dB

4 K Hz

± 3.6 dB

8 K Hz

± 5.6 dB

Remark :

The above spec. are tested under the

environment  RF Field Strength less than

3 V/M & frequency less than 30 MHz

only.

Calibrator

B & K (Bruel & kjaer),  multi-fuction

acoustic calibrator, model : 4226.

Microphone

Electric condenser microphone.

Size of microphone 1/2 inch standard size.

Calibration VR

Build in external calibration VR, easy to

calibrate on 94 dB level by screw driver.

Data Hold

Hold function to freeze the display value.

Operating Temp.

0 to 50   ( 32 to 122   ).

Operating Humidity Less than 80% RH.

Power Supply

Alkaline or Heavy duty type, 9V d.c.

006 P, MN1604(PP3) or equivalent.

Power Consumption Approx. DC 6 mA.

Dimension

250 x 70 x 28 mm ( 9.9 x 2.8 x 1.1 inch ).

Weight

250 g/0.55 LB ( including battery ).

Standard 

Instruction Manual ...................... 1 PC.

Accessories

Optional 

94 dB Sound Calibrator, 

Accessories

Model : SC-941, SC-942.

Hard carrying case, Model : CA-06.

2

All manuals and user guides at all-guides.com

Summary of Contents for SL-4010

Page 1: ...nt Although this METER is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please read the following instructions caref...

Page 2: ...splay 3 3 3 Range upper lower indicator 3 3 4 Range switch 3 3 5 External calibration VR 3 3 6 Power OFF ON Hold switch 3 3 7 Battery compartment Cover 3 4 MEASURING PROCEDURE 4 5 REPLACEMENT of BATTE...

Page 3: ...ong lasting components including a strong light weight ABS plastic housing case Small and light weight design allow one hand operation Low battery indicator High quality with economical cost 2 SPECIFI...

Page 4: ...ulti fuction acoustic calibrator model 4226 Microphone Electric condenser microphone Size of microphone 1 2 inch standard size Calibration VR Build in external calibration VR easy to calibrate on 94 d...

Page 5: ...condenser microphone 3 2 Display 3 3 Range upper lower indicator 3 4 Range switch 3 5 External calibration VR 3 6 Power ON OFF Hold switch 1 Power ON 0 Power OFF 3 7 Battery compartment Cover 3 All ma...

Page 6: ...e Power ON OFF Hold switch 3 6 Fig 1 to the Hold position will freeze the display value If slide the Power ON OFF Hold switch 3 6 Fig 1 to the 1 position then the display will make the measurement con...

Page 7: ...the optional SOUND CALIBRATOR model SC 941 SC 942 or other equivalent sound calibrator Power on the Sound Calibrator plug calibrator output into the Electric condenser micro phone 3 1 Fig 1 of the So...

Page 8: ...hting Tolerance Hz Character IEC 61672 class 2 31 5 39 4 dB 3 5 dB 63 26 2 dB 2 5 dB 125 16 1 dB 2 0 dB 250 8 6 dB 1 9 dB 500 3 2 dB 1 9 dB 1 K 0 dB 1 4 dB 2 K 1 2 dB 2 6 dB 4 K 1 dB 3 6 dB 8 K 1 1 dB...

Reviews: