Active (Real) Power

Range

Resolution

Accuracy

0.000 to 9.999 KW

*0.001/0.01/0.1 KW

± (1%+0.008KW)

10.00 to 99.99 KW

*0.01/0.1 KW

± (1%+0.08KW)

100.0 to 999.9 KW

  0.1 KW

± (1%+0.8KW)

1.000 to 9.999 MW

  0.001 MW

± (1%+0.008MW)

* : The resolution is changed according the different ACA range.

Apparent Power

Range

Resolution

Accuracy

0.000 to 9.999 KVA

*0.001/0.01/0.1KVA

± (1%+0.008KVA)

10.00 to 99.99 KVA

*0.01/0.1 KVA

± (1%+0.08KVA)

100.0 to 999.9 KVA

  0.1 KVA

± (1%+0.8KVA)

1.000 to 9.999 MVA

  0.001 MVA

± (1%+0.008MVA)

* : The resolution is changed according the different ACA range.

Reactive Power

Range

Resolution

Accuracy

0.000 to 9.999 KVAR

*0.001/0.01/0.1KVAR

± (1%+0.008 KVAR)

10.00 to 99.99 KVAR

*0.01/0.1 KVAR

± (1%+0.08 KVAR)

100.0 to 999.9 KVAR

  0.1 KVAR

± (1%+0.8 KVAR)

1.000 to 9.999 MVAR

  0.001 MVAR

± (1%+0.008 MVAR)

* : The resolution is changed according the different ACA range.

6

All manuals and user guides at all-guides.com

Summary of Contents for DW-6093

Page 1: ...o n measurement A l t h o u g h t h i s POWER ANALYZER is a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please read...

Page 2: ...voltage current to the input terminal Remove test leads before open the battery cover Cleaning Only use the dry cloth to clean the plastic case Environment Conditions Installation Categories III 600V...

Page 3: ...3W one phase by three wires measurement 39 5 3 3 3W three phase by three wires measurement 41 5 4 3 4W three phase by four wires measurement 42 5 5 The CT and PT measurement 44 5 6 ZERO adjustment fo...

Page 4: ...00 and PT ratio 1 to 1000 ACV input impedance is 10 Mega ohms Safety Standard IEC 1010 CAT III 600V Built in clock and Calendar real time data record with SD memory card sampling time set from 2 to 72...

Page 5: ...3W 3P 4W connections Voltage ranges 10 ACV to 600 ACV auto range Current probe Current probe input signal volage ACV input signal 200mV 300mV 500mV 1V 2V 3V and range Current probe input current rang...

Page 6: ...ta logger into SD memory card and down load the all the measured value with the time information year month data hour minute second down load to the Excel Integration time for data logger 2 seconds to...

Page 7: ...TL88 4AT 1 Set 4 PCs Alligator clips TL88 4AC 1 Set 4 PCs Clamp Probe CP 1201 3 PCs AC to DC 9V adapter 1 PC SD card 2 G 1 PC Carrying bag 1 PC Optional 2000 Amp current probe CP 2000 Accessories 200...

Page 8: ...1000 A Power factor Range Resolution Accuracy 0 00 to 1 00 0 01 0 04 Remark PFH Long term power factor PF For 3 4W 3 3W PF PF1 PF2 PF3 3 For 1 3W PF PF1 PF2 2 Phase angle Range Resolution Accuracy 180...

Page 9: ...1 0 1KVA 1 0 008KVA 10 00 to 99 99 KVA 0 01 0 1 KVA 1 0 08KVA 100 0 to 999 9 KVA 0 1 KVA 1 0 8KVA 1 000 to 9 999 MVA 0 001 MVA 1 0 008MVA The resolution is changed according the different ACA range Re...

Page 10: ...curacy 0 000 to 9 999 KVAH 0 001 KVAH 2 0 008 KVAH 10 00 to 99 99 KVAH 0 01 KVAH 2 0 08 KVAH 100 0 to 999 9 KVAH 0 1 KVAH 2 0 8 KVAH 1 000 to 9 999 MVAH 0 001 MVAH 2 0 008 MVAH VAR Hour Reactive Power...

Page 11: ...3 FRONT PANEL DESCRIPTION Fig 1 8 All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 12: ...ey button 3 12 Setup key button 3 13 Voltage input terminals 3 14A Current probe signal input sockets 3 14B Current probe power sockets 3 15 SD card socket 3 16 RS232 socket 3 17 Reset button 3 18 DC...

Page 13: ...ill show as SD Check along with blinking while inserting SD CARD then disappears after several seconds that indicates the data from SD CARD has been read completed 2 The bottom right display of screen...

Page 14: ...000KVAH QH 0 000 KVARH FREQ 0 0 Hz CP1201 SD 20A 3 4W SEC 2 CT 1 PT 1 Check screen 1 4 2 V12 0 0 V V1 0 0 V A1 0 00 A V23 0 0 V V2 0 0 V A2 0 00 A V31 0 0 V V3 0 0 V A3 0 00 A P1 0 000 KW S1 0 000KVA...

Page 15: ...KEY 3 5 Fig 1 Press the key to freeze the display reading 6 BACKLIGHT KEY 3 6 Fig 1 Press the key to switch LCD backlight to ON OFF 7 SETUP KEY 3 12 Fig 1 Press the key to setup the function before me...

Page 16: ...Decimal Basic SD Format 0 Clamp Type CP1201 Use Size 388 KB A Range 20A Free Size 1946 MB V Range 200mV Total Size 1946 MB RS232 Out Sel PT 1 1 V1 I1 P1 CT 1 1 S1 Q1 PF1 Beep ON 1 WH FREQ Year Month D...

Page 17: ...t PT Set the potential transformer from 1 to 1000 CT Set the current transformer from 1 to 600 Beep Set to ON OFF for buzzer Clamp Type Select the Clamp Type to CP 1201 CP 200 CP 2000 CP 3000 or Other...

Page 18: ...P1 CT 1 1 S1 Q1 PF1 Beep ON 1 WH FREQ Year Month Date Hour Minute Second 2010 12 05 11 14 49 screen 2 4 5 1 Folder Name WTA01 SETUP File Name 3P401001 XLS SHIFT 1 REC Date 2008 11 28 00 03 17 Samplin...

Page 19: ...ill show NO File indicator in REC Date option when the selected file is new as screen 1 B The screen will show recording date and time in REC Date option when the selected file has been recorded as sc...

Page 20: ...to 050 Remark When press or 2 sec the setting no will change fast 1P201001 1P2 means one phase by two wires 01 means folder number 001 means file number 1P301001 1P3 means one phase by three wires 01...

Page 21: ...88 KB A Range 20A Free Size 1946 MB V Range 200mV Total Size 1946 MB RS232 Out Sel PT 1 1 V1 I1 P1 CT 1 1 S1 Q1 PF1 Beep ON 1 WH FREQ Year Month Date Hour Minute Second 2010 11 13 14 37 25 Folder Name...

Page 22: ...SHIFT 1 REC Date 2008 11 28 00 03 17 Sampling Time 2 Delet File 0 Decimal Basic SD Format 0 Clamp Type CP1201 Use Size 388 KB A Range 20A Free Size 1946 MB V Range 200mV Total Size 1946 MB RS232 Out...

Page 23: ...se press SETUP KEY in N option to return to screen 1 B Press in screen 1 to enter next setting function Delete File SD Format screen 1 4 5 4 Folder Name WTA01 SETUP File Name 3P401001 XLS SHIFT 1 REC...

Page 24: ...onth Date Hour Minute Second 2010 11 13 14 37 25 4 5 5 SD Format Formatting function for SD CARD A The indicator Y or N will appear on right side display of the option while pressing SHIFT KEY continu...

Page 25: ...Beep ON 1 WH FREQ Year Month Date Hour Minute Second 2010 11 13 14 37 25 screen 2 4 5 5 Folder Name WTA01 SETUP File Name 3P401001 XLS SHIFT 1 REC Date 2008 11 28 00 03 17 Sampling Time 2 Delete File...

Page 26: ...to screen 1 then press to enter next setting function PT CT screen 1 4 5 6 Folder Name WTA01 SETUP File Name 3P401001 XLS SHIFT 1 REC Date 2008 11 28 00 03 17 Sampling Time 2 Delete File 0 Decimal Ba...

Page 27: ...Minute Second 2010 11 13 14 37 25 4 5 7 CT Set the Current Transformer A When press SHIFT KEY once the symbol SHIFT1 will disappear as screen 2 at this time press or can adjust to expect CT values th...

Page 28: ...Q1 PF1 Beep ON 1 WH FREQ Year Month Date Hour Minute Second 2010 11 13 14 37 25 screen 2 4 5 7 Folder Name WTA01 SETUP File Name 3P401001 XLS REC Date 2008 11 28 00 03 17 Sampling Time 2 Delete File 0...

Page 29: ...ng function BEEP Decimal type screen 1 4 5 8 Folder Name WTA01 SETUP File Name 3P401001 XLS SHIFT 1 REC Date 2008 11 28 00 03 17 Sampling Time 2 Delete File 0 Decimal Basic SD Format 0 Clamp Type CP12...

Page 30: ...te Second 2010 11 13 14 37 25 4 5 9 Decimal Type set the Decimal type to Basic or Euro The numerical data structure of SD card is default used the as the decimal for example 20 6 1000 53 But in certai...

Page 31: ...r next setting function Decimal type Clamp type screen 1 4 5 9 Folder Name WTA01 SETUP File Name 3P401001 XLS SHIFT 1 REC Date 2008 11 28 00 03 17 Sampling Time 2 Delete File 0 Decimal Basic SD Format...

Page 32: ...10 11 13 14 37 25 4 5 10 Clamp Type set the clamp type to Lutron Clamp or other Clamp A When press SHIFT KEY once the symbol SHIFT1 will be disappeared and show as screen 2 at this time press or to se...

Page 33: ...Q1 PF1 Beep ON 1 WH FREQ Year Month Date Hour Minute Second 2010 11 13 14 37 25 screen 2 4 5 10 Folder Name WTA01 SETUP File Name 3P401001 XLS REC Date 2008 11 28 00 03 17 Sampling Time 2 Delete File...

Page 34: ...A 300A 3000A Attention The meter s A range Current range value should same as the Clamp s current selecting range value B Press SHIFT KEY once again will return to screen 1 then press to enter next se...

Page 35: ...nd 2010 11 13 14 37 25 4 5 12 V range Setting Voltage range Setting A When press SHIFT KEY once the symbol SHIFT1 will be disappeared and show as screen 2 at this time press or to select V range to 20...

Page 36: ...ON 1 WH FREQ Year Month Date Hour Minute Second 2010 11 13 14 37 25 screen 2 4 5 12 Folder Name WTA01 SETUP File Name 3P401001 XLS REC Date 2008 11 28 00 03 17 Sampling Time 2 Delete File 0 Decimal Ba...

Page 37: ...mpleted press SHIFT KEY continuously at least two seconds again will return to screen 1 and show all the selected items at the same time D Press in screen 1 to enter next setting function RS232 Out Se...

Page 38: ...8 I2 19 Q2 30 WH 9 I3 20 Q3 31 SH 10 P1 21 Q 32 QH 11 P2 22 PF1 33 FREQ screen 3 4 5 13 RS232 OUTPUT SELECT 1 V12 12 P3 23 PF2 2 V23 13 P 24 PF3 3 V31 14 S1 25 PF 4 V1 15 S2 26 PFH 5 V2 16 S3 27 1 6...

Page 39: ...tion to adjust the numbers and the setting value will also be saved during the adjusting E In the function of setting second press or to adjust numbers at this point the number of second is at a stand...

Page 40: ...nd 2010 11 13 14 37 25 4 5 15 When all settings are completed press EXIT KEY to return measuring screen 4 5 16 The descriptions about SD CARD memory space A Use Size To show the space data numbers tha...

Page 41: ...on bottom left display of screen 2 B 2 Connect the line voltage L1 Vn Neutral to V1 and N terminals of the instrument B 3 Place the conductor of CP 1201 A1 to A1 as screen 1 B 4 Connect the output of...

Page 42: ...00 S 1 0 000KVA P F H 0 00 Q 1 0 000KVAR 1 0 0 W H 0 000KWH S H 0 000KVAH Q H 0 000KVARH F R E Q 50 1 Hz CP1201 20A 1 2W SEC 2 CT 1 PT 1 5 2 1 3W one phase by three wires measurement A Diagram screen1...

Page 43: ...Connect the outputs of clamp meter CP 1201 A1 CP 1201 A2 to A1 and A2 terminals of the instrument B 5 The related measuring factors will be appeared on display about the instruction of factor please r...

Page 44: ...en 2 B 2 Connect the line voltage L1 L2 and L3 to V1 V2 and V3 terminals of the instrument B 3 Place the conductor of CP 1201 A1 CP 1201 A2 CP 1201 A3 hook to A1 A2 A3 as screen 1 B 4 Connect the outp...

Page 45: ...3 0 00 A P 0 000 KW S 0 000 KVA Q 0 000 KVAR PF 0 00 P F H 0 00 WH 0 000 KWH SH 0 000 KVAH QH 0 000 KVARH FREQ 50 0 Hz CP1201 20A 3 3W SEC 2 CT 1 PT 1 5 4 3 4W three phases by four wires measurement...

Page 46: ...01 A3 to A1 A2 A3 terminals of the instrument B 5 The related measuring factors will be appeared on display about the instruction of factor please refer appendix 1 5 12 page 51 screen 2 5 4 V12 0 0 V...

Page 47: ...nect the line voltage L1 L2 L3 and Vn to V1 V2 V3 and N terminals of the instrument B 3 Place the conductor of CP 1201 A1 CP 1201 A2 CP 1201 A3 hook to A1 A2 A3 as screen 1 B 4 Connect the outputs of...

Page 48: ...3 0 000 KVAR P 0 000 KW S 0 000 KVA Q 0 000 KVAR PF1 0 00 PF2 0 00 PF3 0 00 PF 0 00 PFH 0 00 1 0 0 2 0 0 3 0 0 WH 0 000 KWH SH 0 000 KVAH QH 0 000 KVARH FREQ 0 0 Hz CP1201 20A 3 4W SEC 2 CT 1 PT 1 5 6...

Page 49: ...A Q3 0 000 KVAR P 0 000 KW S 0 000 KVA Q 0 000 KVAR PF1 0 00 PF2 0 00 PF3 0 00 PF 0 00 PFH 0 00 1 0 0 2 0 0 3 0 0 WH 0 000 KWH SH 0 000 KVAH QH 0 000 KVARH FREQ 0 0 Hz CP1201 Change 20A 3 4W SEC 2 CT...

Page 50: ...SH 0 000 KVAH QH 0 000 KVARH FREQ 0 0 Hz CP1201 REC 20A 3 4W SEC 2 CT 1 PT 1 9 screen 2 5 6 B V12 0 0 V V1 0 0 V A1 0 00 A V23 0 0 V V2 0 0 V A2 0 00 A V31 0 0 V V3 0 0 V A3 0 00 A P1 0 000 KW S1 0 00...

Page 51: ...0 000 KW S 0 000 KVA Q 0 000 KVAR PF1 0 00 PF2 0 00 PF3 0 00 PF 0 00 PFH 0 00 1 0 0 2 0 0 3 0 0 WH 0 000 KWH SH 0 000 KVAH QH 0 000 KVARH FREQ 0 0 Hz CP1201 20A 3 4W SEC 2 CT 1 PT 1 HOLD screen 2 5 7...

Page 52: ...range Setting Current range Setting page 31 Remark The function of the A Range Current Range key is available for the Clamp Type A Range V Range setting only screen 1 5 10 same as screen 2 4 5 11 Fold...

Page 53: ...0 A P1 0 000 KW S1 0 000 KVA Q1 0 000 KVAR P2 0 000 KW S2 0 000 KVA Q2 0 000 KVAR P3 0 000 KW S3 0 000 KVA Q3 0 000 KVAR P 0 000 KW S 0 000 KVA Q 0 000 KVAR PF1 0 00 PF2 0 00 PF3 0 00 PF 0 00 PFH 0 00...

Page 54: ...tive Power VAR PF1 PF2 PF3 Power Factor of each phase PF Total Power Factor PFH Long Term Average Power Factor WH SH 1 2 3 Phase Angle of each phase WH Watt Hour SH Apparent Power Hour QH Reactive Pow...

Page 55: ...LOWBAT indicator ref 5 11 page 50 it should change the batteries 2 open the Battery Cover 3 19 Fig 1 away from the instrument and remove the battery 3 Replace with batteries DC 1 5V AA UM 3 battery X...

Page 56: ...ic application A RS232 lead with the following connection will be required to link the instrument with the PC serial input Meter PC 3 5 mm jack plug 9W D Connector Center Pin Pin 4 Ground shield Pin 2...

Page 57: ...B6 KACV C7 MVAR D8 MVAR Hr B7 MACV C8 GVAR D9 GVAR Hr B8 KACA C9 TVAR E0 TVAR Hr B9 MACA D0 MW Hr D10 Polarity 0 Positive 1 Negative D9 Decimal Point DP position from right to the left 0 No DP 1 1 DP...

Page 58: ...er then connect the SD card adapter into the computer 3 Power ON the computer and run the EXCEL software Down load the saving data file for example the file name 3P401001 XLS 1P201001 XLS 1P301001 XLS...

Page 59: ...EXCEL data screen 2 for example EXCEL data screen 3 for example EXCEL data screen 4 for example 56 All manuals and user guides at all guides com...

Page 60: ...EXCEL data screen 5 for example EXCEL data screen 6 for example EXCEL graphic screen 1 for example 57 All manuals and user guides at all guides com...

Page 61: ...EXCEL graphic screen 2 for example EXCEL graphic screen 3 for example 58 All manuals and user guides at all guides com a l l g u i d e s c o m...

Page 62: ...example 9 PATENT The SD card installation for handheld instruments and the SD card data format structure Data to EXCEL file format already patent pending in the following countries U S A CHINA GERMANY...

Page 63: ...10 THE ADDRESS OF AFTER SERVICE CENTER 60 1011 DW6093 All manuals and user guides at all guides com...

Reviews: