Lutron Electronics DW-6063 User Manual Download Page 12

Meter

PC

(9W 'D" Connector)

Center Pin..........................Pin 4

(3.5 mm jack plug)
 Ground/shield.......................Pin 2

 2.2 K
resistor

Pin 5

The 16 digits data stream will be displayed in the
following  format :

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0

Each digit indicates the following status :
D0

End Word

D1 & D8

Display reading, D1 = LSD, D8 = MSD
For example : 

If the display reading is 1234, then D8 to
D1 is : 00001234

D9

Decimal Point(DP), position from right to the
left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP

D10

Polarity
0 = Positive 1 = Negative

D11 & D12 Annunciator for Display

ACV = 50 

 W = 47

 mA = 53

PF = 54

D13

When send the upper display data = 1
When send the lower display data = 2

D14

4

D15

Start Word

9

Summary of Contents for DW-6063

Page 1: ...into the field of precision measurement Although this METER is a complex and delicate instrument its durable structure will allow many years of use if p r o p e r o p e r a t i n g t e c h n i q u e...

Page 2: ...oltage current to the input terminal Remove test leads before open the battery cover Cleaning Only use the dry cloth to clean the plastic case Environment Conditions Installation Categories III 260V P...

Page 3: ...RCE terminal 4 3 8 RS 232 output terminal 4 3 9 DC 9V Power Adapter input socket 4 3 10 Battery Cover Compartment 4 3 11 Stand 4 4 MEASUREING PROCEDURES 5 4 1 Power ON OFF 5 4 2 ACV measurement proced...

Page 4: ...True RMS measurement of ACV ACA Super large LCD Dual display at same time Full function Auto range Memory Record Max Min Data Hold RS232 USB Computer interface Built in low battery indicator Built in...

Page 5: ...et the USB plug Operation 0 to 50 32 to 122 Temperature Operation Less than 80 RH Humidity Power Supply Alkaline or Heavy duty type DC 9V battery 006P MN1604 PP3 or equivalent Power Approx DC 10 mA Co...

Page 6: ...0 1 mA 0 3 0 3 mA ACA 2 A 1 mA 0 3 3 mA Auto range ACV ACA accuracy is test under input signal is sine wave 50 60 Hz ACV ACA frequency response is from 40 to 400 Hz Max input value AC voltage 260V AC...

Page 7: ...y 3 2 POWER Backlight button 3 3 FUNCTION button 3 4 HOLD button 3 5 RECORD button MAX MIN 3 6 LOAD terminal 3 7 POWER SOURCE terminal 3 8 RS 232 output terminal 3 9 DC 9V Power Adapter input socket 3...

Page 8: ...use the test lead measurement unknown AC voltage 5 Read the up display The value indicated corresponds to the position selected If the Display shows it indicates an Out of range measurement 4 3 ACA m...

Page 9: ...nt power switch Let it work then you can read out the POWER CONSUMPTION from the bottom of the LCD display value 9 At the same time Push the FUNCTION button 3 3 Fig 1 once then change function for mea...

Page 10: ...e the memory function continuously c To exit the memory record function just press the REC button for 2 seconds at least The display will revert to the current reading 4 7 Backlight ON OFF 1 When use...

Page 11: ...d periodically 4 Do not leave the instrument exposed to direct heat from the sun for long periods 5 Before removing the battery compartment cover ensure that the instrument is disconnect from any circ...

Page 12: ...h digit indicates the following status D0 End Word D1 D8 Display reading D1 LSD D8 MSD For example If the display reading is 1234 then D8 to D1 is 00001234 D9 Decimal Point DP position from right to t...

Page 13: ...display show it is necessary to replace the battery However in spec measurement may still be made for several hours after LOW BATTERY INDICATOR appears before the instrument become inaccurate 2 Open...