DPX-IQ Operator’s Manual
AP Spine
A-9
Examine Edge Markers
Make sure the edge markers are positioned above the baseline to include only high
density points. Do not include soft tissue points in the analysis region.
Figure A-11 shows an edge marker position that is not correct: the left edge marker
includes unwanted soft tissue (
➊
), and is positioned on the baseline (
➋
).
NOTE: Calcification, fat deposits, and dense substances can cause
unwanted soft tissue points to be shown on the Line Profile.
Figure A-11. Incorrect Edge marker position
Figure A-12 shows correct edge marker position. The edge marker has been moved so
soft tissue is not included in the analysis region.
To change an edge marker, use the keys
%
or
)
to select an edge marker. Use the
arrow keys to position the edge marker.
Figure A-12. Correct Edge marker position
➊
➋
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...