Luminex 200 System User Manual
28
7. Close the left analyzer door.
8. Prime twice.
As required
Fuses
To replace the fuses in either the Luminex 200 analyzer or the
Luminex XYP instrument:
1. Turn off the power switch on the rear of the analyzer or
instrument, then unplug the instrument power cord from the wall
source. Remove the power cord from the analyzer or instrument.
2. With a small, flathead screwdriver, open the module door on the
lower left corner of the back of the analyzer or instrument. See
Figure 9.
Figure 9 Opening the Module Door
3. Remove the red cartridge (use a flathead screwdriver).
4. Check both fuses for damage.
5. Replace damaged fuses with the type specified on the sticker to
the right of the power input module.
Warning:
To avoid serious injury or death by electric shock,
you must turn off the system and unplug it from the wall.
Summary of Contents for Luminex 200
Page 1: ...Luminex 200 System User Manual Lum nex...
Page 9: ...iii Luminex 200 System User Manual Hardware 47 Software 48 xMAP Reagents 49...
Page 10: ......
Page 42: ......
Page 56: ...Luminex 200 System User Manual 46...
Page 60: ...Luminex 200 System User Manual 50...
Page 64: ......