Lumen Dynamics Group X-Cite 120 Series General Maintenance & Troubleshooting Download Page 5

X-Cite

®

 Guide

3.2  General tips to help maximize lamp lifetime 

Each time the lamp is ignited, the stress on the lamp effectively decreases lamp life by  

1. 

5 hours, therefore it is recommended to avoid shutting down the unit during lunch or  

short breaks or between users, if being used in a multi-user lab 

After lamp ignition, avoid turning the lamp off until it has run for a minimum of 20  

2. 

minutes 

Avoid attempting to strike a hot lamp (note: the X-Cite®’s Intelli-Lamp® will prevent this) 

3. 

Avoid unnecessary movement and jarring of the lamp, especially when it is in  

4. 

operation or hot 

When changing a lamp: 

5. 

Handle the lamp only by the ceramic areas  

a. 

Never touch the glass envelope of the bulb (inner stem), the inner surface, or  

b. 

the outer surface of the reflector. If touched, carefully clean the envelope with  

alcohol. Skin oils can etch the glass and cause premature bulb failure 

Wear cotton gloves or powder free latex/nitrile gloves when handling any lamp 

c. 

Ensure that air filters are kept clear of dust and debris – even a partially blocked filter  

6. 

can reduce lamp lifetime to approximately 25% of what is normally expected.  

Summary of Contents for X-Cite 120 Series

Page 1: ...signal is low to begin with or is excited by a low power portion of the X Cite lamp s output spectrum 2 Liquid Light Guides 2 1 Proper handling and installation of liquid light guides Liquid light gui...

Page 2: ...e when the other end is 4 connected to the X Cite unit if the unit is turned on in this condition the cap will overheat melt and or permanently discolour the quartz end of the light guide Do not expos...

Page 3: ...bles can sometimes recover Disconnect the light guide from the X Cite unit and leave the light guide undisturbed on a shelf for 1 2 weeks For this to be effective it is important to catch the bubble w...

Page 4: ...ife If the lower power regions of the output spectra are being used it may be desirable to replace lamps more frequently to maximize signal levels Elapsed Time Hours 0 200 400 600 800 1000 1200 1400 1...

Page 5: ...amp will prevent this 3 Avoid unnecessary movement and jarring of the lamp especially when it is in 4 operation or hot When changing a lamp 5 Handle the lamp only by the ceramic areas a Never touch th...

Page 6: ...air flow between the X Cite and walls or other equipment 2 Do not remove the rubber feet on the X Cite unit or otherwise reduce block the space between the bottom of the unit and bench top This may c...

Page 7: ...res cleaning or replacing if Dust lint debris has accumulated Dust lint debris is blocking the view of the grey mesh filter The filter looks as though it is crumbling 2 It is possible to clean newer f...

Reviews: