Installing the Lamp Module 

The OmniCure

®

 Series 2000 curing system is available configured in two 

different versions.  One version is for standard curing and one version is for 

surface curing. The appropriate type of lamp must be installed depending on 

which configuration your Series 2000 is set.  
1.

 

Be sure the AC power cord is disconnected from the unit. 

2.

 

Remove the screw from the lamp access side panel using the tool 

provided and remove the panel from the unit cover. 

3.

 

Carefully remove the lamp module from its container, holding only the 

ceramic component and glass rim 

4.

 

As illustrated below, position the lamp facing towards the front of the 
unit with the power leads facing towards you.  

 

Caution: 

Do not touch the bulb’s glass envelope or the inside surface 

of the reflector. Skin oils can cause the lamp module to fail 

prematurely. 

 

5.

 

Make sure the middle of the lamp is in position to fit into the spring 
clamp. Slide the lamp until it snaps into the spring clamp. The leading 
edge of the reflector should fit snugly into the lamp holder recess. 

 

 

 

6.

 

Locate the 4-pin Intelli-Lamp sensor connector at the rear of the lamp module 
and connect it to its mate located on the top of the lamp-housing wall. 

       Tip:

 the Intelli-lamp connector will only attach in the correct orientation. If 

you  are having difficulty attaching the connector, try rotating it by 180º. 

 

7.

 

Locate the Power connector with two leads and connect it to its mate located 
on the bottom of the lamp-housing wall. 

        

Tip:

 the Lamp Power connector will only attach in the correct orientation. If 

you  are having difficulty inserting the connector, try rotating it by 180º. 

 

 

 

 

8.

 

It is very important to ensure that the lamp anode cooling fin (bar) 
and lamp power lead, at the front of the lamp are centered within the 
lamp holder cutout.  Rotate the lamp as required. 

9.

 

Replace the lamp housing side panel and tighten the fastening screw. 

Note 1:

 if the lamp housing panel is not secured completely, the lamp will 

not strike and the “

bulb

” message will display when power is turned on to 

the unit. 

Note 2:

 if the wrong type of lamp has been installed in your unit, the 

message “

E1

” will appear on the LED display.  

 

Exposure to Mercury represents a health hazard to humans.

 

When unpacking or   

installing the lamp, always wear protective clothing and a face mask.

 

 

 

 

 

Rim 

*Center 
lamp 
power 
lead as 
shown