EN
13
Data entry into the register is confirmed by a short illu-
mination of the Status lED in red.
Before the required number of correct measurements
(according to Table 1) is achieved, the device displays arithme-
tic mean from the completed measurements. Subsequent me-
asurements are added applying the walking window principle,
i.e. the “oldest” measurement is deleted and replaced
by the most recent one. When values outside the measuring ran-
ge are measured, the device indicates it and correct measure-
ments are counted from the beginning. Default time is set to 1s.
Table 1
Averaging
time
Number
of averaged
measurements
Refreshing processed value
(time of one
measurement)
0.5 s
1
every 0.5 s
1 s
2
every 0.5 s
3 s
6
every 0.5 s
5 s
10
every 0.5 s
10 s
20
every 0.5 s
15 s
30
every 0.5 s
30 s
60
every 0.5 s
1 min
100
every 0.6 s
2 min
100
every 1.21 s
5 min
100
every 3 s
7 min
100
every 4.2 s
12 min
100
every 6 s
15 min
100
every 9.1 s