7. TECHNICAL DATA
INPUTS:
Pt100
(- 200... + 850)
°
C
Pt500
(- 200... + 850)
°
C
Pt1000
(- 200... + 850)
°
C
J (Fe-CuNi)
(- 100... + 1100)
°
C
K (NiCr-NiAl)
(-100... + 1370)
°
C
N (NiCrSi-NiSi)
(- 100... + 1300)
°
C
E (NiCr-CuNi)
(- 100... + 850)
°
C
R (PtRh13-Pt)
(0... + 1760)
°
C
S (PtRh10-Pt)
(0... + 1760)
°
C
T (Cu-CuNi)
(- 50... + 400)
°
C
Resistance measurement
0... 10 k
W
Voltage measurement
±
300 mV,
input resistance > 9 M
W
,
Voltage measurement
±
600 V,
input resistance > 4.2 M
W
Current measurement
±
40 mA,
input resistance < 4
W
Current measurement
±
5 A,
input resistance = 10 m
W
±
10%
Measuring subranges (preserving the class):
Pt100
320
°
C
Pt500
230
°
C
Pt1000
290
°
C
Thermocouple J
350
°
C, 700
°
C
Thermocouple K
450
°
C, 950
°
C
Thermocouple N
550
°
C, 1000
°
C
Thermocouple E
250
°
C, 520
°
C
Resistance
110
W
, 220
W
, 460
W
, 950
W
, 2100
W
, 5000
W
Voltage
19 mV, 35 mV, 75 mV, 155 mV,
5 V, 11 V, 22 V, 45 V, 90 V, 180 V, 360 V
Current:
5 mA, 11 mA, 23 mA, 1.8 A, 3.8 A
Intensity of current flowing through the resistance thermometer: < 400
m
A
Resistance of conductors linking the resistance thermometer
with the meter:
< 20
W
/1 wire
Thermocouple characteristics acc. EN 60584-1.
Resistance thermometer characteristics acc. IEC 751+A1+A2.
All manuals and user guides at all-guides.com
Summary of Contents for NA6
Page 2: ...all gui des c om...
Page 3: ...All manuals and user guides at all guides com...
Page 5: ...All manuals and user guides at all guides com...
Page 8: ...all gui des c om...
Page 14: ...all gui des c om...
Page 20: ...all gui des c om...
Page 26: ...all gui des c om...
Page 32: ...all gui des c om...
Page 38: ...all gui des c om...
Page 44: ...all gui des c om...
Page 50: ...all gui des c om...
Page 56: ...all gui des c om...
Page 62: ...all gui des c om...
Page 68: ...all gui des c om...
Page 74: ...all gui des c om...
Page 75: ...All manuals and user guides at all guides com...
Page 76: ...All manuals and user guides at all guides com...