M310-HT Manual
Operation
•
13
4
Operation
4.1
Getting Started
1.
Allow the M310-HT to warm-up to room temperature for at least
two hours
after
unpacking and prior to applying power to remove any condensation.
2.
Connect the power cable to a V AC service that matches the power requirement
specification noted in the
Technical Data
section and on the
Specification Label
located on the rear of the M310-HT.
3.
Power the M310-HT by turning on the power switch on the front panel. The controller
will go into self test mode, the fan in the rear will start-up and the blackbody will start
to go the temperature setpoint that was last used prior to power down. The upper
display is the current emitter temperature. The lower display is the setpoint
temperature.
4.1.1
Front Panel Push Buttons
PAGE
Engages access to controller parameters. Do not alter any
configurations.
SCROLL
Engages access to menu items underneath the Page menu.
No need to use this button.
DWN/UP
Used to raise and lower the temperature setpoint.
4.2
Changing the Setpoint
To change the temperature setpoint (lower display), Press either the DWN/UP buttons on the
controller. Each press will advance the setpoint 0.1 degree. To advance faster, press and hold for
5 seconds until it advances a few degrees/second. Let go when it is near the desired setpoint.
Press again to bring it to the exact value desired. Note that this setpoint will be retained when
power is turned off.
4.3
PID Tuning Constants
The PID tuning constants have been factory set for optimum performance. Altering these
parameters will void the warranty and will require assistance from LumaSense to restore it.
Blackbody Temperature
Setpoint Temperature