Figure 4.4. The Status Window displaying the Pump Out Flow Check.
The Multi Gas Monitor – INNOVA 1316A-1 is purged with sample gas
by pumping room air from the zero inlet on the rear panel through the
entire flow system.
During the zero calibration purge period, the O
2
sensor’s electrical
zero offset is calculated and stored. The optimum gain is calculated
for each IR channel.
The new calibration constants:
CO
2
, CO, HC, and NO zero calibration constants.
O
2
span calibration constant.
Are calculated and stored in flash memory:
When the zero calibration process is successful, the following actions
are performed:
The red Zero Required in the Status screen, see
, is
changed to show a green Not Required, see