background image

IMPAC pyrometers IS 210 · IGA 210

 

 

 

 

11 

8.2 

with software 

InfraWin

 

(only via optional available USB adapter)

 

InfraWin

 is an adjustment and evaluation software that can 

be used to read out the available parameters of the pyrome-
ter and change them, if required. Additionally the measure-
ment can be observed via graphical surface.  
Using the software requires the use of the LumaSense USB 
adapter. If connected the software recognizes the pyrome-
ter automatically and the preset parameters can be dis-
played and changed via keyboard or a choice box. 
 
 
The following functions are available: 
 

  Reading and changing pyrometer parame-

ters 

  Graphical display of the current temperature 

or the temperature progress 

  Recording of measurement curves 

  Subsequent analysis, graphically or in table 

form, for print-out or export 

  Spot size calculator to calculate  

spot sizes in any distances 

 
 
 
 
 
 
 
 
 

9

 

Transport, packaging, storage 

 
With faulty shipping the instrument can be damaged or destroyed. To transport or store the 
instrument, please use the original box or a box padded with sufficient shock-absorbing mate-
rial. For storage in humid areas or shipment overseas, the device should be placed in welded 
foil (ideally along with silica gel) to protect it from humidity. 

The pyrometer is designed for a storage temperature of -20 … 70°C with non-condensing 
conditions. A storing out of these conditions can damage or malfunction the pyrometer.  
 
 

10

 

Maintenance 

 

10.1 

Safety 

Attention during pyrometer services. Should the pyrometer be integrated in a running machine 
process the machine has to be switched off and secured against restart before servicing the 
pyrometer. 
 
 

Summary of Contents for IGA 210

Page 1: ...IS 210 IGA 210 IMPAC Pyrometer Proven Quality Operation Manual Betriebsanleitung...

Page 2: ...a USB adapter onto a PC 6 3 4 Connection to the portable parametrizing device HT 6000 7 4 Mechanical Installation accessories 7 5 Optics 7 6 Alignment with the LED targeting light 8 7 Parameter Settin...

Page 3: ...s that arise from the use of any exam ples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect LumaSense Technologies reserves the right to...

Page 4: ...peratures 75 C the output signal is set to approx 3 9 mA Load max 700 at 24 V power supply 400 18 V 1000 30 V Service interface for parameter programming with portable setup device HT 6000 or via USB...

Page 5: ...dditional devices operating with this mains voltage e g transformers Touching mains voltage can be fatal An incorrect connection and mounting can cause serious health or material damages Only qualifie...

Page 6: ...corder Digital display 230 V 4 20 mA white brown Schirm schwarz 24 V DC Power supply green yellow Power supply for LED targeting light White 24 V Pin 1 Brown 0 V Pin 2 Black Shield Green LED Pin 5 Yel...

Page 7: ...least as big as the spot size of the pyrometer in the meas uring distance The following drawings show the pyrometer s spot sizes M in different measuring distances a The spot size M in a distance of a...

Page 8: ...ch are different from the nominal distance of the selected optics may be calculated using the following formula D D M a a M 2 2 D D M a a M 1 1 6 Alignment with the LED targeting light For exact measu...

Page 9: ...r the two spectral ranges of the in struments are listed in the emissivity table below The tolerance of the emissivity values for each material is mainly dependent on the surface conditions Rough surf...

Page 10: ...until a new hot object appears in the measuring beam The temperature which has to be recognized as hot is defined by the low limit of the adjusted sub range The stored maximum value will be deleted wh...

Page 11: ...surement curves Subsequent analysis graphically or in table form for print out or export Spot size calculator to calculate spot sizes in any distances 9 Transport packaging storage With faulty shippin...

Page 12: ...suring distance see 5 Emissivity set too high Set lower correct emissivity corresponding to the material see 7 2 Lens contaminated Clean lens carefully see 10 2 Temperature indication too high Emissiv...

Page 13: ...60 Connection cable 5 m 3 821 770 Connection cable 10 m 3 821 780 Connection cable 15 m 3 821 790 Connection cable 20 m 3 821 800 Connection cable 25 m 3 821 810 Connection cable 30 m 3 826 500 Portab...

Reviews: