Model 4404-16 Geo-Exploration System
Technical Manual
Section 3
Ludlum Measurements, Inc.
Page 22
April 2018
Calibration Wizard
The Calibration Wizard guides the user through the process of setting the
calibration constants for each online channel. The following parameters will
be set by the wizard: HV actual calibration constant, HV read back
calibration constant, LLD actual calibration constant, and ULD actual
calibration constant. The wizard automatically calculates the calibration
constant based on the measurements entered in for each channel. All
calibration constants are zeroed out at the beginning. Calibration requires a
Model 500 Pulser with a high voltage readout or a high impedance voltmeter
with at least 1000 Megohm meter input resistance.
HV Actual Calibration Constant
Click the “Begin” button to start the process of calibrating the HV actual
calibration constant. The HV actual calibration constant is calculated
automatically by entering the high voltage measured at the detector