Ludlum Measurements 336641 Manual Download Page 10

 

Model 3A Survey Meter 

 
 

 

 
 

page 3  

rate-plot on semi-log paper. Set the operating 
voltage at the flattest position of the curve near 
the knee of the curve. 

 

 

For GM detectors, adjust the HV to 900 volts. 
For special applications, the power supply may 
be adjusted from 400 to 1500 volts. 
 

Note:

 Measure high voltage with a Model 500 pulser 

or  a  high-impedance  voltmeter  with  a  high-meg 
probe. If one of these instruments is not available, 
use a voltmeter with a minimum of 1000 megohm 
input  resistance.  Do  not  use  a  vacuum  tube  type 

voltmeter  for  this  adjustment  unless  an  external, 
high-voltage multiplier probe is used. 
 

5.3  Alarm Circuit

 

After HV and range controls are set, the alarm set 
control  may  be  adjusted  to  the  desired  alarm 
threshold.  Minimum  setting  is  10%  of  the  meter 
scale;  maximum  setting  is  full-scale  deflection. 
When the alarm circuit is energized, both the audio 
and  visual  indications  should  occur.  Pushing  the 
reset button resets the meter and alarm circuit. 

 
6.   MAINTENANCE

 

Instrument  maintenance  consists  of  keeping  the 
instrument  clean  and  periodically  checking  the 
batteries and the calibration.  
 
To  assure  proper  operation  of  the  instrument 
between  calibrations,  the  instrument  should  be 
tested  with  a  check  source  prior  to  each  use.  A 
reference reading should be obtained when exposed 
to the check source in a constant and reproducible 
manner at the time of calibration. If the instrument 
response differs from the reference reading by more 
than ±20%, the instrument should be returned to a 
calibration  facility  for  maintenance,  repair,  or 
recalibration as required. 
 
Recalibration  should  be  accomplished  after  any 
maintenance  or  adjustment  of  any  kind  has  been 
performed on the instrument.  Battery replacements 

are not considered to be maintenance, and do not 
normally require the instrument to be recalibrated. 
 
Ludlum Measurements recommends recalibration at 
intervals  no  greater  than  one  year.  Check  the 
appropriate  regulatory  agency’s  regulations  to 
determine required recalibration intervals.  
 
The  batteries  should  be  removed  and  the  battery 
contacts  cleaned  of  any  corrosion  at  least  every 
three months. If the instrument has been exposed to 
a  very  dusty  or  corrosive  atmosphere,  service 
batteries more frequently. 
 
Use  a  spanner  wrench  to  unscrew  the  battery 
contact  insulators,  exposing  the  internal  contacts 
and  battery  springs.  Removing  the  handle  will 
facilitate access to these contacts. 

 
 

Note:

 Never store the instrument over 

30  days  without  removing  batteries. 
Although this instrument will operate at 
very  high  ambient  temperatures, 
battery  seal  failure  can  occur  at 
temperatures as low as 38 

˚C (100 ˚F). 

Summary of Contents for 336641

Page 1: ...LUDLUM MODEL 3A SURVEY METER June 2020 Serial No 336641 and Succeeding Serial Numbers ...

Page 2: ...LUDLUM MODEL 3A SURVEY METER June 2020 Serial No 336641 and Succeeding Serial Numbers ...

Page 3: ......

Page 4: ... anyimplied warranty of merchantability or fitness which extend beyond the description of the face there of If the product does not perform as warranted herein purchaser s sole remedy shall be repair or replacement at the option of Ludlum Measurements In no event will Ludlum Measurements be liable for damages lost revenue lost wages or any other incidental or consequential damages arising from the...

Page 5: ......

Page 6: ...ONS 1 4 OPERATING PROCEDURES 2 5 CALIBRATION 2 5 1 Calibrating CPS Scale 2 5 2 Detector Operating Point 2 5 3 Alarm Circuit 3 6 MAINTENANCE 3 7 PARTS LIST 4 Model 3A Survey Meter 4 Main Circuit Board Drawing 464 x 639 4 Calibration Board Drawing 363 x 650 5 Wiring Diagram Drawing 363 x 653 5 DRAWINGS AND DIAGRAMS 7 ...

Page 7: ...Model 3A Survey Meter page iii ...

Page 8: ...ange multipliers of X0 1 X1 Xl0 and X100 THRESHOLD 40 10 mV AUDIO built in unimorph speaker with an ON OFF switch HIGH VOLTAGE externally adjustable from 400 to 1500 volts RESPONSE 4 or 22 seconds 10 90 of final reading LINEARITY 10 of true value ALARM INDICATION audible and visual indication when above alarm threshold ALARM RANGE 0 to off scale for each range multiple latching alarm BATTERY DEPEN...

Page 9: ...Note The enter post of the battery is positive Close the battery box lid Switch the range switch to BAT The meter should deflect to the battery check portion of the meter scale If the meter does not respond recheck that the batteries have proper polarity Connect the cable to the instrument and detector Turn the instrument range switch to X100 Expose the detector to a check source The speaker shoul...

Page 10: ...h a check source prior to each use A reference reading should be obtained when exposed to the check source in a constant and reproducible manner at the time of calibration If the instrument response differs from the reference reading by more than 20 the instrument should be returned to a calibration facility for maintenance repair or recalibration as required Recalibration should be accomplished a...

Page 11: ...5 Q15 MPSW51AG 05 5765 Q44 2N3904G 05 5755 Q96 2N3904G 05 5755 Q145 2N4402BU 05 5763 Q200 2N4402BU 05 5763 Q201 2N7000 05 5820 Q213 2N3904G 05 5755 INTEGRATED CIRCUITS U1 CMXT3904TRLF 05 5888 Ref No Description Part No U2 TLC372 06 6265 U3 CD4098 06 6066 U4 CMXT3906TRLF 05 5890 U5 ICM7555 06 6136 U6 TLC27M7IP 06 6248 U7 MAX631 06 6249 U8 CMXT3906TRLF 05 5890 U9 U11 CMXT3904TRLF 05 5888 U101 LM385Z...

Page 12: ...Ref No Description Part No Wiring Diagram Drawing 363 x 653 AUDIO DS1 UNIMORPH 6030 21 9251 CONNECTORS J1 1 640442 5 MTA100 13 8383 J2 640442 2 MTA100 13 8178 J3 640442 5 MTA100 13 8140 J4 640442 4 MTA100 13 8170 J5 RECPT UG706 U SCREW IN C 13 7751 SWITCHES S1 PA 600 210 08 6501 S2 S3 7101 SYZ QE 08 6511 S4 30 1 PB GRAYHIL 08 6517 BATTERY B1 B2 DURACELL D 21 9313 MISCELLANEOUS DS1 BULB 338 21 9307...

Page 13: ... HARNESS PORT CAN WIRES 8363 462 PORT HANDLE ROLLED W SCREWS 4363 139 PORT HANDLE FOR CLIP W SCREWS 7363 203 REPLACEMENT CABLE STD 1 m 39 in 40 1004 CLIP 44 3 TYPE W SCREWS 4002 026 01 CLIP 44 7 TYPE W SCREWS 4010 007 01 CLIP 44 6 TYPE W SCREWS 4010 008 01 ...

Page 14: ...D DIAGRAMS Main Circuit Board Drawing No 464 x 639 Main Circuit Board Component Layout Drawing 464 x 642 2 sheets Calibration Board Drawing No 363 x 650 Calibration Board Component Layout Drawing 363 x 651 Wiring Diagram Drawing No 363 x 653 ...

Page 15: ......

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ......

Page 20: ......

Reviews: