Model 16 Analyzer
Technical Manual
Section 3
Ludlum Measurements, Inc.
Page 3-1
July 2011
Specifications
0
Compatible Detectors
: GM, proportional, and scintillation
Power
: two
˝
D
˝
cell batteries housed in a sealed externally accessible
compartment
Battery Life
: typically 600 hours with alkaline batteries (battery
condition may be checked on the meter)
Multipliers
: ×1, ×10, ×100, and ×1000 selected by a front panel range
selector switch
High Voltage
: adjustable from 400 to 2500 V; can be read on the meter
face
Threshold
: adjustable from 2 to 60 mV
Window
: adjustable from 0 to 60 mV above threshold (can be turned on
or off)
Overload
: senses detector saturation; indicated by red lamp meter and
full-scale meter deflection (adjustable depending on the detector
selected)
Meter
: 6.4 cm (2.5 in.) arc, 1 mA analog type
Meter Dial
: 0-500 counts per minute (cpm), 0-2.5 kV,
BAT TEST
(others
available)
Range
: typically 0-500,000 cpm
Linearity
: reading within 10% of true value with detector connected
Battery Dependence
: instrument calibration change less than 3%
within battery check limits on the meter
Calibration Controls
: individual potentiometers for each range;
accessible from the front cover while in operational status