LSHTEC 86192 User Manual Download Page 4

5-20-2021      Page 4 of 7 

     

 

 

3. Calibration 

If  the  meter  gets  powered  on  in  calibration  holder,  it  will  enter  the  calibration  interface.  The  user  can 
perform the calibration operation according to the prompts, the instrument will enter measurement interface 
after the calibration. If it is not powered on in the calibration holder, it will skip the calibration and enter the 
measurement interface directly. 

 

If it prompts that the calibration failed, the reasons may include: 

  If the standard is not clean, please clean the standard with special lens cloth before placing the meter 

into calibration holder. 

  If the meter is not clung to the calibration holder, please re-place into calibration holder. 

  If there is a significant change in the ambient temperature, restart the device for measurement after 

the temperature of the instrument is close to the ambient temperature and it is confirmed that there 
is no condensation on the lens in the test port and the standard of the base 

  If the light source cannot work normally due to attenuation, the device needs to be sent back to the 

factory for inspection and maintenance 

When the instrument prompts "Calibration failed”, long press the button can skip the calibration. 

4. Measurement 

(1)  Normal Mode 

If the instrument is set to normal mode, remove the calibration holder after calibration, place the 
measuring port on the surface of the object to measure, the instrument will display the measured value 
instantly. Short press the button, the lower left corner of the interface displays the "HOLD" symbol, and 