
Troubleshooting
5
5.2.2.INTERFEROMETER AND LASER
To achieve the best possible results, the interferometer optics must be clean and
correctly aligned. In some cases, it is possible to detect alignment problems
during setup. The operator should always be extremely observant while leveling
the interferometer, and adjusting beam coincidence and verticality so any
problems in misalignment can be detected and corrected before observations
begin. Table 5-2 gives a list of problems and possible solutions which are related
to the interferometer.
Problem
Solution
WEO laser not producing beam
Check cable connections. If the
front panel show no sign of power
check the fuse and power switches.
WEO laser power too low
Laser optics need cleaning.
Alignment of back laser mirror
should be adjusted.
Iodine cell needs rotation
adjustment to maximize beam.
Laser tube needs to be replaced.
Fiber alignment needs to be
optimized
WEO laser will not lock
Ambient temperature too high
(30C)
Laser not warmed up
Laser sweep turned on
Incorrect toroid voltage setting
Large systematic signal in the least-
squares residual plot.
Interferometer base and dropper
tripod are mechanically coupled.
The top interferometer plates are
not fastened.
5-5
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......