Troubleshooting
5
Large residual amplitude
Check parameter file, do not fit
past point 170
Bad cables, especially PIO ribbon
cable
Bad grounding of fringe signal
10nm residual frequency swept
Wrong laser type selected
Incorrect laser modulation
frequency entered in the
fg5param.dat file
No fringes at all
No power to APD.
One/both beam blockers are
pushed in
Very bad alignment of the
test/reference beams. Look into
the telescope and make sure that
the test and reference beams are
overlapped. Also make sure that
the beams are overlapped in the
fringe viewer with the twiddler
Fringe amplitude varies as the cart is
moved up or down.
The iodine-stabilized laser is
running multi-mode. This can
indicate the laser servo gain being
set too high.
Table 5-2
Interferometer Problems and Solutions
5-7
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......