Digital filter (number of samples)
real value
real value
measured value
with great number
of points in digital
filter
measured value
with low number
of points in digital
filter
measur
ed v
alue
me
7.5 Resoluon
The resoluon allows you to set the increment of the last digits of the measure by a constant:
Field choice: se ng values
-
-
-
-
-
-
1 2 5 10 20 50 100
The
is implemented to increase the stability of the measurement in dynamic
LHM
LR-
Cal
applicaons at the expense of the reading divisions.
7.6 Point Pos.
(decimal point posion)
By default, the decimal point posion adopted in the display is set during the calibraon of the
built-in or external transducer. Changing the unit, the
automacally calculates the best
LHM
LR-
Cal
posion of the decimal point for the current unit.
It is possible to force the posion of the decimal point on this page. Selecng
the automac
AUTO
calculaon is enabled, otherwise the format selected is adopted. With a selecon different from
AUTO,
changing the measurement unit is necessary to set the decimal point posion more
appropriate.
7.7 Parameters
•
First Peak Enable
In some applicaons it is necessary to calculate as peak value than the maximum value
recorded but rather the first occurence of a peak value immediately followed by a sharp
decline in the value.
Subsequently, the value may increase again.
To enable this feature, select
Enable First Peak: YES
In a manner correlated it is necessary to define a threshold parameter (as % with respect to
the full scale value) for discriminang small variaon decreasing the value compared to what is
expected:
Peak Threshold (% FS): XX
Increase or decrease the values using the arrow-keys. The figure on next page describes
graphically the meaning of the paramter. To exit select
and confirm.
Exit