2 2
3
IMPORT
ANT SAFETY INF
ORMA
TION
>
>
INSTALLATION
(CONT.)
HUBER 360 MD
CABLE
FIXED CONNECTION (HUBER360 MD)
ET STANDARD CONNECTION (PROHIBITED)
FIXED
CONNECTION
STANDARD
CONNECTION
INSTALLATION
COMPLETED
The installation of your HUBER 360 MD machine must respect the instructions given in the
unpacking notice (to be found in the pocket affixed to the box). The HUBER 360 MD is a non-
mobile device and must be connected by an approved installation expert in accordance with
all local/national regulations.
• Before any electrical installation work is undertaken, the electrical current must be
disconnected.
• Ensure that a differential device (detecting the DC and AC components from a frequency
inverter) is installed on the circuit upstream from the machine.This must comply strictly
with all appropriate safety regulations and restrictions.
• Ensure that the circuit-breaker rating is in correct correspondence to the HUBER 360 MD
(the upstream circuit breaker must be rated higher than the machine).
Place the wall connection box (with isolating cover) as close to the main electrical supply
as possible for a permanent connection to the HUBER 360 MD. Do not connect with a
standard removable plug and socket.
• Locate the electrical system neutral, ensuring that the voltage on the neutral to ground is
close to 0 V.
• Ensure the correct wiring of the phase, earth and neutral in the HUBER 360 MD.
• When the connection is completed, re-connect the electrical current and start the
HUBER 360 MD by activating the on/off circuit breaker.
• Ensure that the HUBER 360 MD is activated and functioning correctly.
• The unit must be connected to a dedicated single power line.