![LP-RESEARCH LPMS-B2 Hardware Manual Download Page 8](http://html.mh-extra.com/html/lp-research/lpms-b2/lpms-b2_hardware-manual_1943408008.webp)
LPMS-B2 Series Hardware Manual ver. 1.0
- 5 -
Table 4-2. Accelerometer Specification
Parameters
Typical Value
Unit
Measurement range
±2/±4/±8/±16
g
Sensitivity
0.061/0.122/0.244/0.488
mg/LSB
Linear acceleration
sensitivity change
vs. temperature
±1
%
Linear acceleration
typical zero-g level
offset accuracy
±40
mg
Linear acceleration
zero-rate change
vs. temperature
±0.5
mg/°C
Acceleration noise
density
90
(FS= ±2 g ODR = 104 Hz)
μg/√Hz
Table 4-3. Gyroscope Specification
Parameters
Typical Value
Unit
Measurement range
±125/±245/±500/±1000/±2000
dps
Sensitivity
4.375/8.75/17.50/35/70
mdps/LSB
Angular rate
sensitivity change
vs. temperature
±1.5
%
Angular rate typical
zero-rate level
±10
dps
Angular rate typical
zero-rate level
change vs.
temperature
±0.05
dps/°C
Rate noise density
7
mdps/√Hz