5 Settings in the menu
18
|
EN
LM
T
68
54
1a
0
6/
20
21
5.4.5 System menu (device settings)
Information about the parameters in this menu can be found
in the table below.
Device settings submenu
5.5 Configuring and enabling
ventilation programs
You can preconfigure and enable up to four programs for
the patient in the device. If the patient needs different
ventilation settings during the day compared to during the
night, for example, he or she can switch the program him or
herself.
1. Call up the
System
>
Program settings
menu.
2. To copy settings from one program to another, select
Options
.
3. To enable the desired programs for the patient, mark the
Available
button.
Alarm/event list
Lists the alarms and events which have
occurred in chronological order.
Parameter overview
Lists all parameters and set values for
the up to 4 ventilation programs
which can be configured.
Device usage
Obtain information here about the
patient’s therapy (duration, days used,
program proportions) and about
device usage (operating time, internal
battery life remaining or charging
state of internal battery in percent).
PARAMETER
DESCRIPTION
Program settings
You can copy the settings for one
ventilation program to another
program here.
You can enable ventilation programs
for the patient here.
Circuit test
Perform a circuit test here on change
of patient and as required. This
checks for resistance, compliance
and leaks
enabling ventilation programs”, page
18)
.
FiO
2
cell
Activate or deactivate the FiO
2
cell
and calibrate the FiO
2
cell here.
Export therapy data
Here you can export therapy data and
save it to a USB-C flash drive, for
example. A USB flash drive must be
connected for exporting.
Device settings
You can configure the device here
(see “ Device settings submenu”,
page 18)
Device status
Obtain information here about the
device (name, type, serial number of
device and components, firmware
version).
PARAMETER
DESCRIPTION
Alarm volume
You can set the alarm level here.
In doing so, you specify the minimum
alarm volume the patient may set. The
patient can then only set a louder level.
1= very quiet, 2= quiet, 3= loud,
4= very loud
You can test the alarms here.
Display
You can set brightness, orientation,
and the screen background here.
Language
You can set the language and, if
required, another language for the
alarm messages here.
PARAMETER
DESCRIPTION
Service timer
You can activate the reminder function
for the next service here. Select the
period for the next service. Once the
service has been performed, you can
reset the reminder function here.
Filter timer
You can activate and reset the filter
change reminder function here.
Pressure unit
You can set the desired unit of pressure
here.
Exit expert menu
automatically
You can set here whether the expert
menu is exited automatically if there is
no input for an extended period.
Date and time
You can set the current date and time
here.
Delete therapy
data
You can delete therapy data here (in
Standby only).
Set factory settings
Here you can reset the device, including
the alarms, to factory settings (in
Standby only).
Risk of injury from programs being enabled
incorrectly!
Selecting an incorrect or non-configured program may result
in incorrect ventilation settings, putting the patient at risk.
Only enable ventilation programs if they have been
configured for the patient in question.
On change of patient, reset the device to factory
settings.