3 Operation
25
EN
Symbol
Description
Meaning of display
button illuminates continuously green
device is running, vacuum setting “High” is adjusted (� -0.9 bar)
button illuminates continuously red
canister full/filter blocked (motor switches off automatically)
button illuminates continuously green
device is running, vacuum setting “Medium” is adjusted (� -0.5 bar)
button illuminates continuously red
canister full/filter blocked (motor switches off automatically)
button illuminates continuously green
device is running, vacuum setting “Low” is adjusted (� -0.2 bar)
button illuminates continuously red
canister full/filter blocked (motor switches off automatically)
display illuminates continuously green
battery capacity 100 % to 60 %
display illuminates continuously orange
battery capacity < 60 % to 10 %
display illuminates continuously red
Battery capacity < 10 %, connect power supply unit immediately. Switch off the device after
completion of the aspiration procedure for charging!
display blinks green
Battery is being charged.
display blinks red
Overtemperature protection, interruption of charging process until battery temperature returns to permissib-
le range.
Battery defective, operation only possible with connected power supply unit. Please contact your service
partner!
display illuminates continuously green
Power supply unit is connected, voltage available.
display blinks red
Incorrect power supply unit is connected, no charging of battery possible!
button illuminates continuously green
Device is switched on and ready for use.
button illuminates continuously orange
Device still runs for 30 sec. before it switches off automatically. Afterwards the button will still be illuminated
for 10 seconds.
button illuminates continuously red
Internal error, operation is not possible. Please contact your service partner!