background image

Operating Information

Pens

The blue pen has a longer pen arm and records humidity or dew
point (depending on your dip switch selection, see page 6) and the
red pen has a shorter arm and records temperature. The pens are
offset to allow the red pen to glide under the blue pen. The blue
pen indicates the correct time and the red pen precedes it by 3/16˝
of an inch.

The pens move in increments across the chart as sensor readings
change. The display provides smoother and faster readings than
the pens. At any given time there may be a slight discrepancy in
the position of the pen and the reading on the display due to hys-
teresis.

For visual spot checks the display is more accurate than the pen
position but both are within the stated specifications of the units.
(See “Specifications” on page 2)

Pen Adjust

Pen adjustment may be needed if the pens and display do not
match exactly

1. Make sure the pen lifting bar is lowered and press the “Pen

Home” key.

2. Turn the chart hub clockwise rotating the chart.

3. With a small screwdriver loosen the pen adjust screw on the

pen arms and adjust the pen tips to the outer most circle on
the chart.

4. Retighten the screws. After pressing “Pen Home” again the

pens should now read correctly with the display.

Probe

The CR10 probe sits in the cradle in the back of the unit. The CR10
probe comes with the standard 8˝ cord. An extension cord can be

ordered to allow for remote sensing capabilities. (See
“Replacement Parts” on page 8)

Cord Installation

If you have ordered an extension cord for remote sensing capabil-
ities follow the instructions for cord installation listed below:

1. Turn the unit upside down or on its side so that you can see

where the probe connects to the back of the CR10.

2. Twist and pull the black rubber strain relief, beginning on the

back of the recorder body. (See “Instrument Anatomy” on page
4)

3. Slide the strain relief down the cord.

4. Inside you will see a standard connector which looks like a

phone jack. Using a small screwdriver, press retention tab and
the connector will pop out easily.

5. Remove wand from cradle by sliding upward. Repeat steps 1

through 3 for the portion of the cord that connects to the wand.

6. Return the wand to the cradle mounted position by sliding the

wand down into the cradle until it fits into the grips. The probe
can also be replaced by positioning it in back of the cradle area
and pressing it into the cradle until it snaps in place.

Power Supply

We recommend using AC power with four “D” batteries installed as
a back-up power source. This ensures that your recording will not
be interrupted when there is a power failure. When the instrument
is using battery power, the unit will update very slowly to extend
battery life. The 120V AC adapter plugs into the back of the
recorder beneath the probe cradle.

Display Symbols

Alarm is on.

LO BATT

Low battery indicator.

UC

User Calibration mode. This is displayed in the upper right hand
corner of the display.

Reading update indicator. These rectangles will flash along the bot-
tom of the display as long as the recorder is taking readings. If you
are using batteries as the power source the update indicator rec-
tangles will still flash but the rate will be very slow in some modes
and “B” will light on display. The rectangle furthest to the right is lit
continuously when the unit is using battery power.

MIN

This symbol is displayed when you are setting a minimum alarm or
when you have pressed the “Min/Max” key and the minimum value
is being displayed.

MAX

This symbol is displayed when you are setting a maximum alarm or
when you have pressed the “Min/Max” key and the maximum value
is being displayed.

4

E-90-CR10  6/22/05  3:14 PM  Page 4

Summary of Contents for CR10

Page 1: ...Bulletin E 90 CR10 CR10 Temperature Humidity Dew Point Recorder LOVE CONTROLS Division of Dwyer Instruments Inc Michigan City IN U S A 46360 OPERATION MANUAL E 90 CR10 6 22 05 3 14 PM Page 1 ...

Page 2: ...20 to 50 C 40 to 110 F 5 to 40 C Humidity Range 0 to 95 RH non condensing Dew Point 22 to 122 F 30 to 50 C Sensors Temperature Thermistor Humidity Thin Film Capacitor Accuracy Temperature 1 8 F 1 C Humidity 2 RH between 0 and 60 3 RH between 61 and 95 Dew Point 95 to 30 RH 2C 2C 30 to 10 RH 4C 3C 10 to 5 RH 8C 7C 5 to 0 RH 49C 48C Probe 92 diameter x 5 9 long on an 8 cord 2 3cm x 15cm long on a 20...

Page 3: ...Instrument Anatomy 3 E 90 CR10 6 22 05 3 14 PM Page 3 ...

Page 4: ...pull the black rubber strain relief beginning on the back of the recorder body See Instrument Anatomy on page 4 3 Slide the strain relief down the cord 4 Inside you will see a standard connector which looks like a phone jack Using a small screwdriver press retention tab and the connector will pop out easily 5 Remove wand from cradle by sliding upward Repeat steps 1 through 3 for the portion of the...

Page 5: ...tely 1 minute and then it will stop For a visual alarm the variable that has moved outside of the minimum or maximum points that you set will flash i e if you set a maximum alarm for 80 F and the temperature reaches 80 F the tempera ture portion of the display will flash Min Max 1 Press the Min Max key once and the minimum value during the recording time will be displayed for approximately 7 sec o...

Page 6: ...s or change the calibration For more information on use of the keypad see Keypad Operation on page 6 7 Keypad Unlocked 8 Off Keypad Locked 8 On Calibration Your instrument was carefully tested and calibrated before being shipped from the factory Additional calibration is not required However should calibration be desired in the future identify a stan dard that is more accurate than this unit Calib...

Page 7: ...Match the CR10 reading with either 11 or 75 depending on the salt capsule being used To raise the RH Dew Point reading press the Up Arrow key To lower the reading press the Down Arrow key 5 When calibration is complete simply press the On Off key to save calibration settings Calibrate Temperature 1 Press the On Off key Up Arrow key simultaneously to enter the User Calibration mode Use dip switch 3...

Page 8: ... also be hysteresis see p 4 Pens Need for pen adjustment see p 4 Pen Adjust Keypad may be locked check dip switch 8 Slower firmer key presses may take multiple presses Check AC adapter connection widen prongs of adapter to fit tighter Check battery installation reinstall Display may be off check dip switch 7 Probe should be dried out under normal ambient conditions Time required varies on Temp RH ...

Page 9: ...0 will be free from defects in material and workmanship for a period of twelve months after delivery In the event of a claim under this warranty the prod uct or part must be returned to the factory for repair or replace ment shipping prepaid with a Return Authorization Number See Factory Service Returns information above It will be repaired or replaced after evaluation The foregoing warranty and r...

Reviews: