LORD AP5103 Series Technical Manual Download Page 8

VERIFICATION OF NORMAL OPERATION

It is the purpose of this section to aid the user in determining, in the event of a malfunction of which

the Series AP5103 is suspected of contributing, whether the instrument is functioning normally or

whether it is the source of the observed trouble. In the event the unit requires repair, the user may

also contact the factory Service Department or the local Daytronic Representative for assistance. 

Daytronic service information is located on the last page of this document.

If the instrument is suspected of faulty operation, perform the following steps.

A. 

If the unit is totally inoperational (front-panel power indicator does not 

light), check the primary power input terminals for proper connection. 

Input power can be from 11 to 28 Vdc and will draw less than 2 watts. 

If properly connected, the front panel Green LED will be illuminated. Before 

reapplying power, visually inspect the power supply and the input power 

connections for any discrepancy which could have caused the overload.

B. 

If the transducer has some preloading, the BALANCE controls may not 

allow successful zeroing of the instrument output. This condition can be 

remedied by connecting a resistor (50 k- 200 k range, metal-film type) 

from the +Signal terminal of the transducer to the + or –Excitation Sense 

terminals. The Excitation terminal to which the connection is made is 

determined by the direction of the loading or off-zero reading.

C. 

The inability to balance correctly, where the instrument output reads 

totally off scale and the BALANCE controls have no authority, can very 

likely be the result of a damaged or defective transducer or cable. This 

possibility can be confirmed (or eliminated) by substituting a transducer 

and cable known to be in good condition or by simulating a balanced 

transducer, using either a commercially available transducer simulator 

or the simple star bridge arrangement shown in Figure 7. The star bridge 

simulates a conventional four-arm bridge in an exact condition of bal- 

ance. To construct a star bridge, connect four 10% carbon resistors as 

shown in Figure 5. Use 180-ohm resistors to simulate a 350-ohm bridge 

and use 56-ohm resistors to simulate a 120-ohm bridge. Neither the 

resistor values nor temperature characteristics are critical since the 

balance condition of a star bridge is not determined by the resistance 

values. Solder two resistors together, then solder the remaining two 

resistors together. Next, connect the two junctions together using a 

separate wire as shown. There is a good reason for this method of 

construction, and it should be followed. Connect the substitute or simu- 

l

ated transducer to the instrument I/O connector using a short 4-wire 

cable configuration as shown in Figure 4. Attempt to balance the substitute 

simulated transducer. If conditions now appear to be normal, the transducer 

or cable is at fault. If the previous difficulties persist, the AP5103 may be faulty.

8 of 9

Summary of Contents for AP5103 Series

Page 1: ...All Rights Reserved Part No 234200B DCN 10087 237 Commerce Drive Amherst NY 14228 USA Tel 716 250 1900 Fax 716 250 1909 Web stellartech com Email info stellartech com Due to the nature of technology...

Page 2: ...tion technique which is outlined in section three The supplied calibration resistor is 59 kilohms 1 which is located internally to the AP5103 case Series AP5103 DC Strain Gage Module Access switch set...

Page 3: ...ent damage Common Mode Rejection Ratio 60 dB 1 2 excitation Input Impedance Differential and Common Mode 10 000 M Offset user adjustable vs Temperature 30 ppm C vs Time 10 ppm month Gain Accuracy Limi...

Page 4: ...e recommended transducer cabling would be eight wire individually shielded twisted pair wired as indicated Fig 1 Sense lines must be connected at the transducer as recommended or at the AP5103 screw t...

Page 5: ...e transducer calibration data mentions some other resistance value In Section 4 of this manual the techniques described above are demonstrated If however the installed value of calibration resistor is...

Page 6: ...to obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Page 7: ...mote Calibration Check The instrument can be placed in the calibration mode positive or negative by shorting pin Power Common and SHP for positive or SHN for negative on the lower I O connector Figure...

Page 8: ...is determined by the direction of the loading or off zero reading C The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authority c...

Page 9: ...ance will be made for any expenses incurred for correcting any defective workmanship and or material without written consent by Stellar Technology This warranty is in lieu of all other warranties expr...

Reviews: