System Operation Overview
22
adjustments to sampling settings to find the balance of sample rate, burst duration and burst
intervals most suitable for the application.
Figure 11 -
Synchronized Sampling Settings
The following considerations should be made when determining sampling settings:
l
Maximum burst length is determined by sample rate. The higher the sample rate, the
shorter the burst lengths can be (
Table 3 - Sample Rate and Burst Duration
).
Sample rate
(kHz)
Maximum burst
duration (seconds)
1
150
10
15
12.5
12
25
6
50
3
62.5
2.4
78
1.92
104
1.4
Table 3 -
Sample Rate and Burst Duration