![LORAD AFFINITY Series Service Manual Download Page 147](http://html.mh-extra.com/html/lorad/affinity-series/affinity-series_service-manual_1939281147.webp)
P/N 9-500-0255
6-19
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
Gantry Components—Remove and Replace
2.5.4
Low Voltage Power Supply
The Low Voltage Power Supply is mounted to the right side Gantry frame, below the
Interface/Fuse Board.
1. Remove power to unit and access the Low Voltage Power Supply by removing
the right side Gantry cover as per Section 2.1.1.
2. Remove the wire harness connectors from their jacks at P1 and P2.
3. Remove the 4 screws from the standoffs on the Gantry frame. Note that a
ground wire is secured to the board using the lower right side screw.
4. Install the replacement Low Voltage Power Supply by reversing this procedure.
Be sure to secure the ground wire using the lower right side mounting screw.
5. Using a DMM, verify the +5 V as follows:
A. On the Host Microprocessor Board, connect the negative lead to TP10
(AGND), connect the positive lead to TP18 (+5 V).
B. On the DMM you should read +5.10 V (+.01 V), if not, adjust the +5 V
potentiometer on the LVPS.
2.5.5
C-arm Safety Microprocessor Board
The C-arm Safety Microprocessor Board is mounted to the right side Gantry frame,
near the bottom.
1. Remove power to unit and access the C-arm Safety Microprocessor Board by
removing the right side Gantry cover as per Section 2.1.1.
2. Remove the wire harness connectors from their jacks at RJ1, RJ2, RJ3 and RJ5.
3. Lift the board off of the 4 “squeeze” type standoffs.
4. Reverse these steps to install the replacement C-arm Safety Microprocessor
Board. Do not replace the Gantry right side cover until after the board is
calibrated.
5. Perform the C-arm Safety Microprocessor Board Calibration procedures as per
Chapter 4, Section 8.4, before releasing the unit to the user.
Note…
The replacement board must be calibrated before the unit is release to the
user.