6-4
P/N 9-500-0255
Service Manual
Chapter 6: Maintenance—Remove and Replace Procedures
Gantry Components—Remove and Replace
2.1.1
Left and Right Side Panels
Removal of the Gantry’s left and right side panels (Figure 6-2, Items 1 and 2) provide
access to the unit’s high voltage electronics, power distribution, and microprocessor
assemblies. Both cover removal procedures are identical except when the radiation
shield is installed. The radiation shield must be removed before removing the right-
side gantry cover. If radiation shield not installed perform steps 1 and 4 through 8. If
shield is installed perform Steps 1 and 4 through 8 for left side covers. Perform all
steps for right side covers.
1. Remove power to unit.
2. Remove hardware securing the radiation shield lower bracket to the bottom of
the Gantry.
3. Remove hardware securing the Gantry upper bracket to the top of the radiation
shield. Remove shield and set in a secure location.
4. Remove hardware securing top of side cover.
5. Slide the cover upward until the tabs on the cover clear the mounting slots in
the Gantry frame.
6. Pull the cover slightly away from the unit, then set it upright on the floor. Note
that a wiring harness connected to the Angle Display Board and the Emergency
Off Switch is still attached to the Gantry. Proceed to Step 8 if side panel is to be
replaced.
7. Do not perform steps 8 and 9 if the unit is to be powered on at any point while
servicing the unit. The unit will not function if the cables/side panels are
disconnected.
8. Turn the cover away from the unit to gain access to the wiring harness. Follow
the harness to the connector on the Gantry frame.
9. Disconnect the wiring harness, then set the cover in a secure location.
2.1.2
Upper Rear Gantry Panel
The upper rear panel (Figure 6-2, Item 3) on the Gantry provides access to the Power
On/Off Circuit Breaker, and the components on the C-arm’s Vertical Travel Assembly
(VTA).
1. Remove power to unit.
2. Loosen the eight fasteners that secure the upper rear panel to the Gantry.
3. Lift the panel up slightly, then pull it away from the unit.
4. Store the panel in a safe location to prevent damage.