Marina Shower stretcher |
Instructions for use
|
39
Franç
ais
Deutsch
English
Nederlands
4. Safety functions
4.1. Lock button
(Only for electrically adjustable versions)
The lock button is located in the middle of the undercarriage.
Once the lock button has been depressed all electrical functio-
ning will stop. Rotate the lock button a quarter of a turn to the
right to unlock it
(See figure 12)
.
The orange LED light on the hand control will flash
when the Lock button has been activated.
4.2. Sleep setting
(only for electrically adjustable versions)
After around an hour, the stretcher will fall into a ‘sleep’ setting and the battery indication
LEDs on the manual control will go out. In order to use the stretcher again, it must be
activated
(See paragraph 2.2 Activating).
4.3. Emergency lowering
(only for electrically adjustable versions)
If the controls stop working and the stretcher is in a raised
position, it can be lowered using the emergency lowering
button
(See figure 13).
Check first to make sure the stretcher is not in sleep setting!
• Remove the plug from the manual control (1).
• Press the emergency lowering button (2).
The stretcher will now return to the lowest position (make sure no persons or objects are
under the stretcher). Take the client from the stretcher and make sure the device is not
used until it has been repaired. Place the stretcher in a safe place and indicate that it is
out of order.
Figure 12
Lock button.
Figuur 13
Emergency lowering button.
1
2
Summary of Contents for Marina 6100 2300