4.2. Sleep setting
After around an hour, the chair will fall into a ‘sleep’ setting and the battery indication
LEDs on the manual control will go out . In order to use the chair again, it must be
activated
(See paragraph 2.2. Activating).
4.3. Emergency lowering
If the controls stop working and the chair is in a raised
position, it can be lowered using the emergency lowering
button .
Check first to make sure the chair is not in sleep setting!
• Remove the plug from the manual control (1)
(See
figure 25)
.
• Press the emergency lowering button (2)
(See figure 25)
.
The chair will now return to the lowest position (make sure no persons or objects are
under the chair) . Take the client from the chair and make sure the device is not used
until it has been repaired . Place the chair in a safe place and indicate that it is out of
order .
Contact the technical service or Lopital’s service
department . E-mail: info@lopital .nl
4.4. On / Off switch
On the right hand side of the cover cap, there is the on /
off switch
(See figure 26).
If the chair is out of use for an
extended period, you are advised to turn it off . This will
prevent any unauthorised use of the chair .
5. Cleaning and maintenance
The Elexo must be cleaned after every use, with water and regular household cleaning
agents . PUR integrated foam components with IMC coating should be rubbed clean and
disinfected . Do not use any abrasive or other aggressive cleaning products . If necessary,
the PUR components can be disinfected with a solution of 70% ethanol or propanol .
Figure 25 Emergency lowering
button .
2