LONGER BT100-2J Operating Manual Download Page 2

1

2

S1

S2

P3

Prime Key

P4

P2

Speed Decrease Key

OP1

P1

Direction Key

Start/Stop Key

Press the Start/Stop Key to start or stop the pump. This key is
invalid in prime state when pump runs at full speed.

Direction Key

Press the Direction Key to change the rotating direction of the
pump and the Direction Indicator lights at the same time. This
key is invalid in prime state.

Speed Increase/decrease Key

Press the Speed Increase Key or Decrease Key one time to
increase or decrease the speed in step of 0.1 rpm. Continuously
press the increase key or decrease key to change the speed
rapidly. The Speed Increase Key and Decrease Key are invalid in
prime state.

Prime Key

In common state, press the Prime Key to enter prime state when
the pump runs at full speed (that is 100 rpm) and the LED displays
“---” for empting, filling and rinsing operation; Press the Prime
Key again to return to common state.

LED Display

Speed Increase Key

Start/Stop Key

Introduction

Acceptable pump heads for BT100-2J are YZ1515x, YZ2515x, YZII15,
YZII25, DG-1 and DG-2. The pump delivers flow rates from 0.002 to
380 mL/min. The speed can be adjusted manually or automatically
through external control interface. Store the running parameter
automatically. Easy to operate.

S

Standard Configuration

S1 Pump Head
S2 Drive

OP

Optional Fittings

OP1 External Control Module

Reference:

Make sure the power line of the pump is grounded reliably to

protect the operator in humid condition.

For detail information about external control function please see

page 4.

P

Parts

P1

Power Switch

P2
P3
P4

External Control Interface

Power Socket
Build-in Fuse

Basic Operation

Direction
Indicator

External Control
Indicator

BT100-2J OPERATING MANUAL

BT100-2J OPERATING MANUAL