Direct Calibration - Zero Setting
How to calibrate the
ZERO
point.
You can set Zero first, if you prefer, but you will not be able to change the decimal point
position in the ZERO calibration step.
This will not be an issue if your zero calibration reading is 0, but may become
confusing otherwise.
When you have finished your calibration, please remember to put the calibration
lockout switch in its ON position, to protect your settings.
OK
Max/Min
Reset
Lockout Switch must be OFF
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Apply the lowest calibration signal
you can achieve, ideally 0% of
system capacity.
Press OK.
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
You can now set the display value you
want to see - use DIGIT to select each
digit in turn, and the UP or DOWN
arrow to increase or decrease each
digit’s value, as required.
Set1
Set2
Output
Alarms
Digit
OK
Max/Min
Reset
Press to accept
1
2
3
4
Done!
Display shows
direCt
followed by
set Lo
OFF
Circuit board ON
Set1
Set2
Output
Alarms
Digit
Press 3 seconds