How to adjust your Analogue Output
The lockout switch should be set ‘OFF’ to change the analogue output calibration.
You can set the analogue output range to suit your display range.
The analogue output can be directly proportional or inversely proportional to
the display range, for example you can have 4-20mA output for display 0 to100
or for display 100 to 0.
1)
Press ‘OUTPUT’ button for 3 seconds
2)
0% O/P LED should flash
3)
Set the display for the reading value where you want 0% Output, by using
DIGIT and UP/DOWN buttons
4)
When set, press the OK button.
5)
Now the 100% O/P LED should flash
6)
Set the display to the reading value where you want 100% Output, by using
DIGIT and UP/DOWN buttons
7)
When set, press OK, to complete your adjustment of the analogue output
scaling
Please remeber to set the lockout switch ‘ON’ to save your settings.
Page 15
DIGIT
Min.
ZERO
SPAN / dp
ZERO
SPAN & d.p.
MAX / MIN
TARE
ALARMS
OK
s
RESET
s
OUTPUT
AL1
AL2
AL3
AL4
0% O/P 100% O/P
Limeariser
Max.
DIGIT
Min.
ZERO
SPAN / dp
ZERO
SPAN & d.p.
MAX / MIN
TARE
ALARMS
OK
s
RESET
s
OUTPUT
AL1
AL2
AL3
AL4
0% O/P 100% O/P
Lineariser
Max.
Jumpers on Analogue Board
0% gives
4-20mA
4mA
0-10V
0 V
-5 to +5V
-5V
So in this example, if you set the jumpers for
4-20mA, you will get 4mA output when the
display is 50.00
Jumpers on Analogue Board 100% gives
4-20mA
20mA
0-10V
10 V
-5 to +5V
+5V
So in this example, if you set the jumpers for
4-20mA you will get 20mA output when the
display is 350.00