background image

29

Error  codes  and  fault  finding

--Ur--
--Or--

1. Under Range. The meter is being asked to display a value which is more
negative than its limit of -199999

2. Over Range. The meter is being asked to display a value which is higher
than its limit of 999999

If no loadcell is connected to the display, you will see one of these fault
codes.

In addition, these fault codes could be displayed because the signal is too
negative, too positive, or because there is a wiring error, or because the
display’s scaling has been adjusted to give excessive sensitivity, or because
there is a fault in the display.

a. Measure the input signal on terminals 3(-) and 4(+)and confirm that it is
between  -40mV DC and +40mV DC. You can temporarily  force the dis-
play’s input signal  to 0mV by wiring a link between terminal 3 and terminal 4,
but keep the loadcell connected in addition to this link, as the display needs
to see a full bridge at its input

b. Measure the excitation voltage on terminals 1(-) and 6(+) and confirm that
it is between 9.5 and 10.5V DC. If it is less than 9.5V, remove the plug-in
connector and measure the voltage directly on pins 1 and 6 with the connec-
tor removed. If it is still less than 9.5V, and if the meter has not been pur-
posefully modified to give lower excitation, then there is most likely a fault
with the display and it should be returned to us for the attention of our QA
Manager, with a report of what you found, for repair.

c. Measure the sense voltage on terminals 2(-) and 5(+) and confirm that it is
between 4 and 10.5V DC. If it is not, please confirm that you have connected
terminal 1 to terminal 2 and terminal 5 to terminal 6, either at the meter end
of the loadcell cable or at the loadcell end.

d. Measure the voltage between terminal 2(-) and terminal 3(+) It should be
half of the voltage you measured in step 3. If it is not, there is most likely a
wiring error between the display and the loadcell, or the loadcell is faulty.
Check the wiring according to the installation instructions for the loadcell and
the display. If that appears to be correct, measure the resistance of the four
loadcell branches according to the loadcell manufacturer’s specifications.

e. If steps 1 to 4 were satisfactory, the problem may be caused by the scal-
ing being too sensitive. If you are happy to do so, it is a simple matter to reset
the display’s scaling to the factory default conditions. Put the lockout switch
off and press all 4 buttons together for around 3 seconds until the display
shows “dEFS n” press the UP button once so the display shows “dEFS Y”.
Now press the OK button and the display will be reset to factory defaults.

If still no success, please return the display to us for the attention of our QA
Manager, with a report of what you found, for repair.

Summary of Contents for IL2-L

Page 1: ...nd general information Caution There is a risk of electrical shock if this instrument is not properly installed Caution Risk of danger Read the whole manual before you install this meter London Electr...

Page 2: ...2...

Page 3: ...Calibration Modes 17 Direct Calibration 18 19 Theoretical Calibration 20 21 Sensor Drift correction 22 Logic Input connections Front Buttons 23 Factory defaults 24 CalibrationAudit number 24 Signal F...

Page 4: ...rs fromelectric shock We designed this meter for Pollution Degree 2 environments only Power supplies to this equipment must have anti surge T fuses rated at 400mA for 230V supply 400mA for 110V supply...

Page 5: ...hutdown periods you may have to to wait several weeks for an equivalent replacement or we may have no stock at the time you urgently need it You mayalso needto pay extra carriagecharges ifyou wanta fa...

Page 6: ...240 VAC or 11 30VDC These meters are designed to mount into a protective enclosure which will protect users from contact with power and signal wiring Theseunits mustbe installedfully assembled and mus...

Page 7: ...of meter U Bracket Bracket screw 1 of 2 Left side of meter case U Bracket Bracket screws U Bracket Bracket screws Slide the meter with gasket into the panel cutout Securing into the panel Re fit the...

Page 8: ...mple using the 4 buttons Output functions include isolated 16 bit 4 20mA 0 10V 10 to 10V RS232 RS485 and up to 4 alarms with a wide choice of alarm functions Install the meters in a suitable protectiv...

Page 9: ...sealed IP65 all round and is supplied with the meter in stalled for you to wire to via 3 cable glands It comes complete with SPC4 splashproof cover We also offer a dual version of this enclosure which...

Page 10: ...5mm2 Tighten screw terminals to 7 0 lb in torque 0 8 Nm torque Cable insulation voltage rating Cables shall have an insulation voltage rating of at least 380V continuous Wire preparation We recommend...

Page 11: ...nly Link Exc to Sense and link Exc to Sense Set Sig to 5V using 2 x 10K resistors Class II Installation Not used Cable and or zener barrier resistance must not cause the excitation voltage to drop bel...

Page 12: ...ngst the twisted pairs can actually worsen noise performance 4 The cable should be routed away from noisy wiring and devices such as power feeds frominverters discharge lighting cables welder cabling...

Page 13: ...13 4 Wire Loadcell 6 Wire Loadcell...

Page 14: ...ing note to tell you whether it is available in Easy orAdvanced mode How to choose menu mode Lockout Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press brie...

Page 15: ...witch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Display shows UI Eng Default for User Interface English or UI FrA for User Interface French 1 2 3 Set1 Set2 Outp...

Page 16: ...ess of the UP button will select a newbrightness level There are 7 brightness levels to choose from Default Full brightness Set1 Set2 Output Alarms Digit OK Max Min Reset Press to accept 1 2 3 Done Di...

Page 17: ...2 Theoretical Calibration this is when you type in the sensor s theoretical signal level at the bottom and top of itsrange and then type in the value the display should show for each signal level How...

Page 18: ...et1 Set2 Output Alarms Digit OK Max Min Reset Press OK Apply the highest calibration signal you can achieve ideally 100 of system capacity You can use less but you will get better accuracy with higher...

Page 19: ...OK Max Min Reset Lockout Switch must be OFF Set1 Set2 Output Alarms Digit OK Max Min Reset Press OK Apply the lowest calibration signal you can achieve ideally 0 of system capacity Set1 Set2 Output Al...

Page 20: ...ch digit s value as required eg set 20 0000 if your high input is 20mV Set1 Set2 Output Alarms Digit OK Max Min Reset Press to accept Set1 Set2 Output Alarms Digit OK Max Min Reset Display now shows r...

Page 21: ...ng Use DIGIT to select each digit in turn and the UP or DOWN arrow to increase or decrease each digit s value as re quired eg set 000 000 if your display range is 0 to 100 000 Theoretical Calibration...

Page 22: ...ings less than 8 least significant counts The corrected value is stored in non volatile memory so that any drift trend will be re ap plied after you have switched off the meter Lockout Switch must be...

Page 23: ...lines to be galvanically isolated from the input signal To enable the front panel buttons you must connect the logic input for that function to common If you want to have both front panel and remote t...

Page 24: ...one This feature is available in Easy and Advanced Modes Display shows defs n Defaults no Press the DOWN button to change the display to defs y Defaults Yes if you want to return to default conditions...

Page 25: ...layed value they will respond at the same rate as the filtered display Lockout Switch must be OFF Press 3 seconds Set1 Set2 Output Alarms Digit OK Max Min Reset Display shows FIL XX X where XX X is th...

Page 26: ...d up to the nearest 1 no rounding 2 5 10 20 or 50 Lockout Switch must be OFF Press 3 seconds Set1 Set2 Output Alarms Digit OK Max Min Reset Press repeatedly until you see LSt X where X is either 1 2 5...

Page 27: ...if you wish once you become more familiar with the setup method Lockout Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press repeatedly until you see dLAY XX...

Page 28: ...Switch must be OFF Press together briefly Set1 Set2 Output Alarms Digit OK Max Min Reset Press OK button briefly and repeatedly until you see boot 0 or boot 1 or boot 2 displayed Set1 Set2 Output Alar...

Page 29: ...most likely a fault with the display and it should be returned to us for the attention of our QA Manager with a report of what you found for repair c Measure the sense voltage on terminals 2 and 5 an...

Page 30: ...o dispose of it It must not be thrown away with general rubbish If you are unable to dispose of this item locally you may send it to us for recycling Conditions 1 We will only accept items of our manu...

Page 31: ...ignals bipolar 4 or 6 wire loadcell up to 40mV 100mV option Input Resistance 10 Megohms Accuracy 0 05 of range Span tempco 25 ppm Degree Celsius Zero Tempco 30 ppm Degree Celsius 5V Excitation voltage...

Page 32: ...0 Version F0 18 Software released Manual format revised to improve clarity and segregate easy from advanced menu functions Optional outputs now described in their own dedicated manuals DIN Rail mounti...

Page 33: ...33 Notes...

Page 34: ...34 Notes...

Page 35: ...35 Notes...

Page 36: ...ncludes relay output wiring All signal cabling shall be screened The screen shall only be terminated to the power earth terminal at the meter end of the cable Declared as true and correct for and on b...

Reviews: