The display will show
In 01
followed by
an editing screen. Here you can enter your
next calibration input signal level using
DIGIT button to select each digit in turn.
Increase or decrease the value of each
digit using the UP or DOWN buttons.
Press OK when set.
The display will now show
rd 01
You can now tell the meter what it should
display for this input signal.
Select each digit in turn with the DIGIT
button, and increase or decrease each
digit’s value using the UP or DOWN
button, until the display is set as required.
Press OK when done.
The display will show
Add.Pt
Press OK if you want to Add another
calibration point, you will repeat the previ-
ous steps, each time incrementing
In XX
and
rd XX
up to In 09 and rd 09 which
are the last available points.
If you want to finish adding points, press
Set2 until the display shows
SAVE
, and
then confirm by pressing OK
If you want to abort the setup, press Set2
until the display shows
qUIt
and then
press OK
Linearisation with Theoretical Cal. - cont’d.
Input
Display
In00
In01
In02
In03
In04
In05
In06
In07
In08
rd 08
rd 07
rd 06
rd 05
rd 04
rd 03
rd 02
rd 01
rd 00
Done!