London Electronics EasyReader ER4P Instruction Manual Download Page 8

Direct 4-20mA I/P

Connections

We supply detachable screw terminal connectors to make installation as easy as possible for you.
You should use multistrand insulated wire with ferrules to DIN46228/1.  You can use stripped wire
with cross sectional area from 0.5 to 2.5mm

2

.  Strip back insulation 7mm.

3 cable glands accept the 3 groups of cables up to 6mm diameter.

Use screened cable for the input signal and connect the screen to power earth at the meter end of
the cable only. For best performance, keep the signal cable well away from the power and alarm

cables, which could carry electrical noise likely to interfere with your measurement.

Input Connection Examples:

+

Loop powered 2 wire
4-20mA transmitter

I/P

4-20mA from
active source.

Direct 0-10V I/P

0-10V from
active source

5

Common

Calibration mode

Tare

Peak/Valley

Reset

Screen

termination

point

Exci

0-10V   +

4-20mA +

Common

Lockout

Exci

0-10V +

4-20mA +

Common

Lockout

Earth

Neutral

Line

Power

Logic I/Ps

Signal I/P

Locked

Unlocked

Locked

Unlocked

Locked

Unlocked

Exci

0-10V +

4-20mA +

Common

Lockout

Exci

0-10V +

4-20mA +

Common

Lockout

You can  operate the Tare,
Peak/Valley and Reset functions
remotely with contact closures,
or with the front panel buttons.
To activate the front panel
buttons, you must make
permanent connections
between the desired function
connector  and Common

Summary of Contents for EasyReader ER4P

Page 1: ...ation Rear case screws please note The rear panel is held in place with finger screws which only need to be gently tightened Do not use tools to tighten or loosen the screws as this could cause damage...

Page 2: ......

Page 3: ...t position selection 7 8 11 Declaration of Conformity 17 Filter time constant selection 7 8 Front Panel controls 11 General Description 3 Getting Started 4 Introduction 1 Input connection examples 5 L...

Page 4: ...ent replacement Or we may have no stock at the time you urgently need it You may also need to pay extra carriage charges if you want a fast guaranteed courier service Warranty repairs or replacements...

Page 5: ...r 110V supply or 1A for DC supplies in the range 11 30VDC Check that the model number and supply voltage suit your application before you install the meter Don t touch any circuitry after you have con...

Page 6: ...the meter in engineering units to give up to 1999 or 9 999 display range with selectable decimal point position You can either scale the meter theoretically using stored calibration figures or direct...

Page 7: ...cables through the cable glands Connect the signal and power cables to the appropriate screw terminal connectors See our connection drawing to check that you are using the correct terminals Check bef...

Page 8: ...arry electrical noise likely to interfere with your measurement Input Connection Examples Loop powered 2 wire 4 20mA transmitter I P 4 20mA from active source Direct 0 10V I P 0 10V from active source...

Page 9: ...e meter either from a calibrator or a sensor in the system This is known as the DIRECT calibration method Signals must be stable Connect the calibration mode terminal to common to enable this method T...

Page 10: ...been selected When correct press the OK button 2 How to set ZERO Press the ZERO button for 3 seconds You ll see In LO briefly then a value with one digit brighter than the other This is 4 000 for a 4...

Page 11: ...ow you to move the decimal point with the UP DOWN buttons 4 When correctly set press OK 2 How to calibrate Zero 1 Apply the Lowest cal signal and ensure it is steady before proceeding 2 Press the ZERO...

Page 12: ...ESETinput Tare Command Press the front panel tare switch for more than 2 seconds or make a momentary contact closure to the TARE terminal and Common if you want to force the display to 0 Following rea...

Page 13: ...MODE A contact closure which selects either direct or theoretical calibration method Contact open Theoretical calibration Contact closed Direct calibration TARE Contact closure resets the displayed v...

Page 14: ...e unlocked ZERO Press for more than 3 seconds Lets you adjust the meter zero calibration SPAN d p Press for more than 3 seconds Lets you adjust the meter span calibration decimal point position filter...

Page 15: ...eight 102mm A 57mm high model is also available Display Rangability Fully digitally scalable for ZERO and SPAN readings anywhere in the range 1999 to 9999 with selectable decimal point positon Viewing...

Page 16: ...if the wall material is not suitable to take screws directly Route cables neatly away from the display If the cables may become damaged in the environment protect them with suitable conduit or trunkin...

Page 17: ...ou may need to use screw plugs or other screw accessories if the material is not suitable to take screws directly Mount the brackets first then fit the display to the brackets Route cables neatly away...

Page 18: ...Record of Revisions 16 December 2003 Product released 25 March 2004 Added Wiring and Jumper record sheet to page 16 8 November 2005 Added side view for wall mounting case on page 13 15...

Page 19: ...16 Wiring and jumper record For future reference record all the jumper positions and wiring colours you used in your installation...

Page 20: ...rs covered by this certificate must be installed in adherence to the following conditions Signal cabling shall be routed separately to power carrying cabling includes relay output wiring All signal ca...

Reviews: