© Copyright 2016 by Lohman Technologies, LLC – All Rights Reserved
Page
23
of
24
Guidance and manufacturers declaration – electromagnetic emissions
The AfibAlert
®
is intended for use in the electromagnetic environment specified below. The customer or the user of the
AfibAlert
®
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The AfibAlert
®
uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The AfibAlert
®
is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not
Applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Not Applicable
Guidance and manufacturers declaration – electromagnetic immunity
The AfibAlert
®
is intended for use in the electromagnetic environment specified below. The customer or the user of the
AfibAlert
®
should assure that it is used in such an environment.
Immunity Test
IEC 60601
Compliance
Electromagnetic
Test level
level
environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
(full compliance)
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%. If the display or unit
does become locked, the unit should be re-started and
a repeat test should be done.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not Applicable
Not Applicable
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
Not Applicable
Not Applicable
Voltage dips, short
interruptions and voltage
variations on power supply
lines
IEC 61000-4-11
<5%
U
t (>95%
dip in
U
t ) for 0,5 cycle
40%
U
t (60% dip in
U
t)
for 5 cycles 70%
U
t
(30% dip
in
U
t) for 25 cycles
<5%
U
t(>95% dip in
U
t)
for 5 sec
Not Applicable
Not Applicable
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
(full compliance)
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
NOTE
U
t is the a.c. mains voltage prior to application of the test level