ONSTAGE INCUBATION SYSTEM
INSTALLATION GUIDE
This Installation Guide provides instructions for installing the onstage incubation system for the CELENA
®
S.
For detailed instructions, refer to the provided user manuals.
VC1706-01
© 2017 Logos Biosystems, Inc. All Rights Reserved.
Logos Biosystems | www.logosbio.com
GAS MIXER
TEMPERATURE CONTROLLER
HUMIDIFYING COLUMN
AIR PRESSURE GENERATOR
HEATED LID AND PLATE
Use a wash bottle to fill the humidifying
column. Don’t exceed 130 mL. Don’t let
the water fall below 50 mL during use.
Don’t leave any input ports open in the
gas mixer. Close unnecessary ports with
dead-end tubing.
Turn on the gas mixer only after the
input pressure of all gases have been
set to 1 bar.
MAX - 130 mL
MIN - 50 mL
After placing your sample onto the
CELENA S, wait 20 minutes before
starting a time-lapse experiment.
Only use the provided tubes, cables,
and plugs to ensure electrical safety.
IMPORTANT POINTS
MAIN COMPONENTS
HEATING INSERTS