Logitech Electronics 2000 MKII Series User Manual Download Page 8

Programming the Instrument

 

Calculating a RATE and a TOTAL Factor 

The value displayed for 

RATE

 is calculated as : 

RATE

 = N x Scaling Value    

N = number of pulses counted in time T 

T x FACTOR 

   

T = measurement time (usually 1 sec) 

The value displayed for 

TOTAL

 is calculated as : 

TOTAL

 = C x Scaling Value   

C = total count. 

FACTOR 

Programmable Parameters 

The following values must be programmed into the unit in order to display the desired values: 

 

Factor   

 

ii 

Total - scaling value   

 

iii

 

Total - decimal point position 

 

iv 

Rate - scaling value 

 

Rate - decimal point position 

Other parameters that may need to be programmed are: 

 

vi 

Low Alarm limit  

 

vii 

High alarm limit 

And if the analogue output option has been specified: 

 

viii

  

Reference 

 

 

ix

  

Full Scale 

In order to program the parameters the instrument must be operating in the 

SET

 mode.   Unless the 

appropriate parameters have been specified at the time of ordering, all values will be factory set to 
zero when the instrument is first powered up.   

Since  zero  is  an  invalid  number  for  Factor  or  Scaling  values,  the  instrument  must  have  a  link 
connected between M0 (terminal 7) and 0V (terminal 9 Common), and will initially be forced into the 
setup routine and will display   

SET 

….

 FACTOR 

….

 0

   (

....

 signifies a one second delay before the 

display changes).   

2000CT-

MKII

_hb v3.1  01/2020 

 

 

 

 

!

8

Summary of Contents for 2000 MKII Series

Page 1: ...2000 MKII SERIES Counter Tachometer Model 2000CT User Guide...

Page 2: ...arameters 8 FACTOR 9 TOTAL Scaling Value 9 TOTAL Decimal Point Position 10 RATE Scaling Value 10 RATE Decimal Point Position 10 LOW Alarm 10 HIGH Alarm 10 Analogue Output Settings if O P module ordere...

Page 3: ...al Signal Inputs Signal B count up Control Inputs Factor protection external Reset Control inputs Opto isolated requires sink to 0V of 220W maximum Outputs DC Voltage 10V nominal unregulated 80mA max...

Page 4: ...of the total display and the rate display so that rate may be displayed as units per second units per minute units per hour milli units per second etc and total can be displayed as units milli units...

Page 5: ...black Connect the brown wire to the terminal marked with the letter L or coloured red The mains supply to this instrument must be protected with a 1 Amp fuse Transformer Connections 2000 MKII SERIES...

Page 6: ...RATE in which case the display will show NO SIG DP Decimal Point Switch This is located under the FUNC switch and is used to set the position of the decimal point as required A single depression of t...

Page 7: ...hown on the display is equal to or greater than the value programmed into memory The LOW alarm operates on a value equal to or below the programmed value The alarm outputs automatically reset when the...

Page 8: ...ion Other parameters that may need to be programmed are vi Low Alarm limit vii High alarm limit And if the analogue output option has been specified viii Reference ix Full Scale In order to program th...

Page 9: ...ues that have been programmed prior to returning the instrument to the RUN mode Otherwise when the required values have been programmed press the RUN SET button once to switch the instrument to the RU...

Page 10: ...ace to the left on each depression of the lower right hand button This sets the position of the decimal point in the display when RATE is being displayed see iii above vi Low Alarm ALA L When the calc...

Page 11: ...from rate display to total display The unit will display TOTAL for one second followed by the current total count Note RATE is displayed using a maximum of six digits TOTAL is displayed using up to th...

Page 12: ...reset prior to using the instrument with the revised programming to avoid the possibility of an incorrectly displayed value which has been held in memory from use with the previous settings Protection...

Page 13: ...ams Front Panel Layout Note On instruments with the SINGLE switch set option only the FUNC switch is accessible unless the bezel and front panel are removed Rear panel layout 2000CT MKII_hb v3 1 01 20...

Page 14: ...Connection Diagrams Typical Applications Variable reluctance Magnetic Pickups Single input operation Dual input operation 2000CT MKII_hb v3 1 01 2020 14...

Page 15: ...Alarm Output Connections Alarm connections with external power supply 2000CT MKII_hb v3 1 01 2020 15...

Page 16: ...advisable to use a good quality mains filter mounted as close to the instrument as possible ensuring that the cable between the filter and the instrument is kept separate from any cables carrying hig...

Page 17: ...he warranty only extends to the original user Copyright Logitech Electronics Limited 1994 to 2017 Document No 2000CT MKII_hb v3 0 July 2017 2000CT MKII_hb v3 1 01 2020 17 Design and Manufacture of Ele...

Reviews: