3.1 Utility and Space Requirements
Electrical: 115 VAC, 5 amps, 60 Hz.
Overall Dimensions: 17.27 inches high x 9.88 inches wide x 5.88 inches deep.
3.2 Installing the UV Bulb
Lamp should be handled only when cool and with clean cotton gloves. Do Not touch the lamp
with bare hands. Oils on your hands will etch the quartz glass, create hot spots, and can reduce
lamp life.
The unit shipped without the UV bulb installed. See instructions below for initial bulb installation.
1. Disconnect the electric power from unit.
2. Remove the three phillips head screws holding the filter assembly to the lamp housing.
3. Remove the filter assembly.
4. Wearing clean cotton gloves, screw the new bulb securely into socket.
5. Replace the filter assembly and screws.
3.3 Replacing the UV Bulb
1. Disconnect the electric power from unit.
2. Allow the lamp to cool.
3. Remove the three phillips head screws holding the filter assembly to the lamp housing.
4. Remove the filter assembly.
5. Wearing clean cotton gloves, remove the old bulb.
6. Screw the new bulb securely into socket.
7. Replace the filter assembly and screws.
3
3
Installation
4
Operation
1. Connect the plug to a 115 volt, 60 Hz. grounded outlet.
2. Move the power switch to the “ON” position. Allow 3 to 5 minutes for the lamp to reach
full intensity.
3. If the lamp will be used intermittently, at frequent intervals, leave the lamp on continuously
rather than turn it on and off. This procedure gives longer bulb life and avoids the lamp
restarting delay.
4. If the lamp is turned off and then back on, it will not restart until the mercury vapor in the bulb
cools. This will take approximately 4 minutes.
5. The lamp head can be positioned on either of the two mounting posts on the power supply or
it can be removed for other mounting or hand held use.
WARNING!
Lamp Housing and Filter will get hot during continuous operation. Take care not to touch these
parts. DO NOT operate the lamp with the filter removed.