21
7.5
Basic Troubleshooting
Malfunction
Possible Cause
Corrective
No display
appears on the
screen
Main power cable is
disconnected
Check that the main power cable is
connected to an AC source.
EMO button has been activated
Turn the EMO button clockwise
until it pops out.
Main switch is not turned on
Turn on the main switch, which is
located on back panel.
Fuse break
Check the fuse in power inlet.
Cannot start
agitation
Lid is opened
Close lid.
Lid switch is malfunction.
Check lid switch and replace it if
defective.
Contact Henkel service
“Lift Timeout”
Elevator cannot move up/down
Manually operate elevator to check
if can run or not:
1.
Yes, please refer next table.
2.
No, please contact Henkel
service.
Limit Switch malfunction
Refer 7.3, find limit switch and
check them. Replace them if
necessary.
8
Care and Maintenance
Warning!
Maintenance procedures should be performed by trained personnel only.