Connect Sensor to Interface 

 

 

§

 

Connect a 4-wire load cell as shown in wiring diagram above. Slots 5 and 8 (nor marked) are to 
be used for calibration only and are not used during normal operation. Connect only the shield 
to slot 8. If you want to connect the voltage output into your own DAQ then you can tap into the 
output signal between pins 5 and 8. 

§

 

If your unit has been shipped with a calibrated load cell, simply plug the provided terminal in to 
the 8-pin terminal receptacle. 

§

 

Connect the included power supply as shown. 

§

 

Connect the included USB cable to the PC as shown. 

 

 

 

Summary of Contents for DI-1000UHS-10K

Page 1: ... acquisition Plug and play operation with LV 1000HS 10K software and factory calibrated load cells Select data acquisition rate from 1KHz to 50Khz True force measurement not just G values Analog calibration feature for easy calibration of your load cell Package Contents DI 1000UHS 10K High Speed USB Load Cell Interface Power supply 12 V The unit will accept between 9 to 24 V USB Cable to connect t...

Page 2: ...are Installation Install the LV 1000Hs 10K software and drivers first before connecting the unit to the computer You MUST reboot when asked to do so by the driver installation program Please note that once you install the LV 1000HS 10K software you must connect the sensor to the DI 1000UHS 10K and then connect the USB cable to the PC You must run InstaCal prior to launching and using the LV 1000HS...

Page 3: ... during normal operation Connect only the shield to slot 8 If you want to connect the voltage output into your own DAQ then you can tap into the output signal between pins 5 and 8 If your unit has been shipped with a calibrated load cell simply plug the provided terminal in to the 8 pin terminal receptacle Connect the included power supply as shown Connect the included USB cable to the PC as shown...

Page 4: ...nit back to us for a certified calibration However if you cannot send it back or you must calibrate it yourself then you can apply known loads to the load cell preferably full load and calibrate the DI 1000UHS 10K interface as follows 1 With no load on the load cell adjust the R1 offset potentiometer pot to read approximately 0 5 V between pins 5 and 8 2 With full load adjust the R2 pot to read ap...

Page 5: ...and environments such as LabView Java NET languages Android etc You can find the specifications API and programming assistance for BTH 1208LS at Measurement Computing web site Note Pin 8 is used to connect and ground the shield of the load cell if one is present as well as to read the output voltage during calibration Please connect a wire if necessary to pin 8 to read the voltage between pins 5 a...

Reviews: