OPERATING MANUAL 

                                                   SONIMIX3030-PERM

1T1PT

Edition 1.0E   January 2003

 Page 1

OMSX3030-PERM1T1PT.1oc   10.01.2003

SONIMIX 3030-PERM1T1PT

OPERATING MANUAL

               1)   INTRODUCTION

               2)   CHARACTERISTICS

                  2.1   GENERAL CHARACTERISTICS

                  2.2   PRESENTATION OF THE INSTRUMENT

                  2.3   INTERNAL STRUCTURES

                  2.4   ELECTRONICS

               3)   SOFTWARE DESCRIPTION AND ACCESSIBLE FUNCTIONS

                  3.1   INTRODUCTION

                  3.2   SOFTWARE EXPLORATION AND DEFINITION OF THE PRINCIPAL KEYS

                  3.3   DETAILS OF THE MENUS AND FUNCTIONS ACCESSIBLE

               4)    INSTALLATION

4.1  INSTALLATION

              5)    MAINTENANCE AND RECALIBRATION

                  5.1 MASS FLOW GENERATOR

                  5.2 PERMEATION TUBES

                  5.3 ON SITE VERIFICATIONS               

                  5.4 REPAIR

                  5.5 FACTORY RECALIBRATION

6)   TROUBLE SHOOTING

               7)    NOTES

                                                                                                                                              

Summary of Contents for 670210300

Page 1: ...HE INSTRUMENT 2 3 INTERNAL STRUCTURES 2 4 ELECTRONICS 3 SOFTWARE DESCRIPTION AND ACCESSIBLE FUNCTIONS 3 1 INTRODUCTION 3 2 SOFTWARE EXPLORATION AND DEFINITION OF THE PRINCIPAL KEYS 3 3 DETAILS OF THE...

Page 2: ...ts MFC sonic Nozzles pressure regulators permeation oven and ozone generator have been developed by LN INDUSTRIES SA Following features have been in the centre of the development High precision and st...

Page 3: ...mbar l s checked by He detector Casing 19 3U 500mm SX 3030 Perm Mpt No d article Option Descriptif 670210300 SONIMIX 3030 Perm Mpt Module perm ation multi points avec 1 MFC bo tier 19 670210302 SONIM...

Page 4: ...bration In Local mode the interface allows to Configure the calibrator and to perform the manual operations of calibrations necessary after the maintenance process on the monitor The major parameters...

Page 5: ...ubes 30 g m 3 Benzen 30 g m 3 Toluen 30 g m 3 O xylen 30 g m 3 M xylen 30 g m 3 P xylen 30 g m 3 Ethyl Benzen Total outflow of mixture 1000 Nml min Permeation tubes protection flow 100 Nml min Stabili...

Page 6: ...rding to CE recommandations Power consumption 110V 60Hz or 230V 50Hz 60Hz 150W Gas connections by 1 4 Swagelok fittings Dimentions and weight 19 3HE 84TE 500mm deep About 18 Kg 2 2 PRESENTATION OF THE...

Page 7: ...rotected by two 1A fuses with a sticker indicating the voltage 2 2 3 Rear facewith Option1 The rear face is composed of 1 the Outlet of mixture fitting Swagelok PVDF1 4 to be connected to analyzers 2...

Page 8: ...EV6 EV7 EV8 EV9 FILTER EV10 RES1 RES2 FILTER FILTER PS1 PS2 RP1 RP2 DRYER 2 3 2 FLUIDIC MODULES The fluidic circuit of the SONIMIX 6000D1 shows four modules the mass flow generator based on a high pre...

Page 9: ...tration elastomer seals are made of high grade Viton b Permeation oven module all metallic part in contact with the high concentration mixtures are made of non reactive corrosion free material such as...

Page 10: ...ards 1 2 and 3 are fixed together on an L shape aluminia cooler located close to the display The boards 4 and 5 are fixed on the front face The three main boards MB3000 PS3000 and FOUR P6L2 and eventu...

Page 11: ...After cold booting or a RESET RST the instrument identify it self during 5 seconds and shows S O N I M I X 6 0 0 0 D 1 B T X C A L I B R A T O R and jumps to the SAMPLE status S A M P L E In order to...

Page 12: ...w CST2 100 Ncc The informations are accessible by selecting first the Features menu and after with the and key to read the data 3 3 2 SETTINGS The SETTINGS menu is intended to consult or modify the de...

Page 13: ...N G S P E R M E A T I O N T U B E 1 Press ENTER to select the right tube 2 for example S E L E C T I O N G A S B E N Z E N E Press on or to chose the gas contained in the tube 2 and then Enter to sele...

Page 14: ...ample PT3 Press on ENTER P T 2 T O L U E N E A M O U N T 1 9 p p b The stored value in our example 19 ppb is issued of a measurement made with a monitor and compared with a reference mixture at a simi...

Page 15: ...election by ENTER M A N U A L C O N T R O L S A M P L E M A N U A L C O N T R O L Z E R O M A N U A L C O N T R O L S P A N When SAMPLE is selected S A M P L E 0 0 2 3 The display shows the function e...

Page 16: ...O 0 0 5 7 S P A N 0 2 3 6 IMPORTANT NOTICE At any time during a zero or span phase the functioning parameters assembled in the BOARD PANEL menu could be monitored by pressing ENTER To exit from the B...

Page 17: ...H E A T T E M P E R A T U R E T M I N 4 4 5 C T P H E A T T E M P E R A T U R E T M A X 4 5 5 C Press MENU to reach the next active parameter 3 3 5 3 P Tube calibration values B O A R D P A N E L S T...

Page 18: ...ruption in the wiring or a damaged board 3 3 6 2 AIR Alarm The alimentation of Air pressure from the Zero Air Generator or from a gas cylinder is not setted at a correct value It is inferior to the mi...

Page 19: ...AIR supply with a stainless steel tube specially processed for chromatography use In addition the pressure reducers fixed on the cylinders must be of good quality and cleaned made of stainless steel...

Page 20: ...ures superior to 5 bar even temporarly may damage definitively the internal pressure regulators 4 1 2 REMOTE COMMAND CONNECTIONS The remote control is made through a digital interface with only 3 sign...

Page 21: ...OPERATING MANUAL SONIMIX3030 PERM 1T1PT Edition 1 0E January 2003 Page 21 OMSX3030 PERM1T1PT 1oc 10 01 2003...

Page 22: ...ERMEATION OVENS TUBES The permeation oven is equiped with high quality long life permeation tubes their life time is expected of about 12 months After 8 to 10 months certain permeation tubes could sho...

Page 23: ...e between 2 5 and 3bar relative Verify the stabilized AIR pressure at the pressure regulator RP1 s outlet If one of the pressures is inferior to the limit values verify the alimenting pressures and in...

Page 24: ...24 OMSX3030 PERM1T1PT 1oc 10 01 2003 5 3 5 1 Processor board MB 3000 5 3 5 2 Power supply PS3000A TEST POINT or NAME VALUE Vcc 5VDC U2 Uo U1 Vcc GND U2 Uo U1 Vcc GND PROCESSOR INPUT ALARMS PS1 AND PS2...

Page 25: ...10 01 2003 GND 0 V V 16VDC without load 12 VDC on load V 8VDC without load 6 VDC on load Vb COMMON 5 3 5 3 KEYBOARD The 18 keys soft keyboard 5 3 5 4 VFD Display The 2X20 characters Vacuum Fluorescen...

Page 26: ...cessary to maintain the integrity of the product and the validity of the warranty 5 5 FACTORY RECALIBRATION The SONIMIX 6000D1 is a reference intrument which was carefully calibrated with high precisi...

Page 27: ...quashed 1 Verify the consumption of your monitor during calibration and zero phases 2 Verify the tightness of the circuit 3 Verify all the fittings and tubings between the calibrator and the monitors...

Page 28: ...iode connection sense No message on the display when starting the instrument 1 One fuse or both are blowed 1 Change the fuse s If the fuse fails repeatedly the problem is probably more important Verif...

Page 29: ...OPERATING MANUAL SONIMIX3030 PERM 1T1PT Edition 1 0E January 2003 Page 29 OMSX3030 PERM1T1PT 1oc 10 01 2003 NOTES Following...