LMS 100NP Programmable Instruction Manual Download Page 10

 

Section 3       PROGRAMMER 

 

INDEX 

Function overview  

B.1 

Getting started  

B.2 

Program run mode  

B.3 

Display functions  

B.4 

Example program  

B.5 

Function map  

B.6 

Function list  

B.7 

Memory allocation table    B.8 

Memory full indicator  

B.9 

Programming example  

B.10 

Program edit example  

B.11 

 

3.0 FUNCTION OVERVIEW 

The Programmer function in Level P enables the instrument to control applications  
needing 

Setpoint

 changes over time. Examples of this are 

Ramp

 changes where a gradual  

Rate

 of change can be set, or 

Step

 changes which are instantaneous. These can be  

separated by 

Soak

 periods during which the process is held at a constant value. Each  

individual time interval of the program or 

Segment

, together with it’s associated moving  

setpoint value can be stored as a unique 

Program

 and for example be represented by the  

diagram below. 
 

Setpoint 

 
 

Soak  

Step 

 
 
 
 
 
Ramp 

 
 
 
 
 
Time (Segments) 

In addition to those settings that determine the segment profile, it is also necessary to set 

program start

 values, together with the preferred 

ramp rate time units

 for each individual  

program. 

At the end of a sequence, a Program can be arranged to repeat (

Loop

), either a specified 

number of 

Cycles

, or continuously. Only one 

Loop

 can be included in a 

Program

. When the 

program is running, the 

Display

 indicates progress through the sequence of segments, and 

can additionally be interrogated for further segment information. 

It is also possible to CALL an already existing program as a sub program that can be 
inserted as a segment of another program. 

To speed up 

Program

 configuration, several 

Edit

 functions have been provided so that 

individual 

Segments

 and 

Programs

 may be 

Deleted

 or 

Inserted

, and an entire 

Program 

may be 

Copied

 and then 

Pasted

 into another that it will replace. 

For safety reasons, three modes of recovery from a power failure are available. These either 
automatically 

Re-start

 the Program from the beginning, 

Continue

 it from where it stopped, or 

Hold

 it waiting for a user re-start. 

Either one or both of the two auxiliary outputs can be configured as 

Event 

outputs.  

Engaging the 

Holdback

 feature will temporarily halt Setpoint ramping to allow the process  

temperature to catch up should it deviate by more than a pre-set amount during a 

Ramp  

segment. 

To afford maximum programming flexibility, memory is allocated dynamically, and not pre- 
allocated. This allows the user the freedom to configure a small number of long programs  
or a larger number of shorter ones, up to the permitted maximum of 126 Segments per 
program, and a limit of 31 Programs. Should these limits be exceeded, or the Programmer  
memory become fully used, the display will read 

ProG FULL

. Programs can be planned  

using the 

Memory Allocation Table

 which details the memory requirements of individual  

segment types. During configuration a check can be kept on memory usage by  
interrogating the 

USEd 

feature of the display to give an instant reading of ‘percentage  

memory used’. 

Finally, once a program has been configured, it can be run from the 

run off/on/hold  

controls in Level P, and in addition a quick access 

run/hold

 toggle is directly available from 

the front panel. 

The Programmer 

Functions List

 describes the full range of available 

Settings

 for each  

Programmer Function 

together with their display mnemonic. The instrument

 

is supplied  

with a suite of 

Factory Settings

 for each 

Function

. These are shown in bold type. 

The Functions Map illustrates the relationship between the 

Functions

 and their 

Settings 

and provides a guide to the 

Keying Operations

 required to navigate around the menu 

when configuring or running a Program.  

10 of 31

                                                                        01/12

Summary of Contents for 100NP

Page 1: ...Cooled Incubators Series 1A 2 3 4 Programmable Instruction Manual 1 of 31 01 12...

Page 2: ...0 Function Overview 10 3 1 Getting Started 11 3 2 Program Run Mode 11 3 3 Display Functions 11 3 4 Example Program 12 3 5 Programmer Function Map 13 14 3 6 Function List Level P Programmer 15 3 7 Mem...

Page 3: ...pacity 280NP 272 litre capacity Series 2 Models 210NP 135 litre capacity 220NP 200 litre capacity 230NP 290 litre capacity 240NP 420 litre capacity Series 3 Models 100NP 90 litre capacity 200NP 227 li...

Page 4: ...o provide greater flexibility for individual use making it suitable for a wide range of applications including Study of plants and insects Tissue Culture Seed germination and vernalisation Sample stor...

Page 5: ...thing Point Electrical Hazard Warning Mains Electrical Supply Within Exhibited where an internal power socket is fitted Refrigeration Isolation Switch Auto Defrost On Off Switch Weee Label Serial Numb...

Page 6: ...an occur Transportation Series 1A 2 If these cabinets have to be carried it is recommended that at least 2 people assist and that gloves should be used on all occasions The cabinet should be kept upri...

Page 7: ...els in Series 1A 2 and 3 is of sheet steel finished in white stoved enamel Models in Series 1A have an internal liner of white pre coated Aluminium The internal liner of models in Series 2 is of a hig...

Page 8: ...so that approximately 100mm of free space exists around the cooled incubator 2 1 7 Adjust levelling foot or feet depending on model so that the cooled incubator is sited firmly Lockable castors where...

Page 9: ...ogramming instructions please refer to Section 3 or use our quick programming guide at the end of Section 3 2 4 INDICATOR CONTROL LAYOUT Series 1A 2 and 3 The Red safety neon only illuminates when an...

Page 10: ...en Pasted into another that it will replace For safety reasons three modes of recovery from a power failure are available These either automatically Re start the Program from the beginning Continue it...

Page 11: ...s Qt and hold for 3 seconds to hold the program Press Qt again and hold for 3 seconds to run the program Note Level P is read only while a program is active To run a stopped program again press and ho...

Page 12: ...unction Interval 45 mins Step Function Target setpoint 85 deg TYPE TYPE TYPE SPRR PCYC SPRR 55 1 105 T SP T SP 30 137 TYPE TYPE TYPE SOAK STEP SPR SINT T SP SPRR 45 85 55 T SP 30 Ramp Function Rate 55...

Page 13: ...rogram 3 now appears Hold the in list program at current position PROG 2 Start program when on selected This locks other options out EDIT SURE Q Q s DEL NO 3 secs Hold the program at RUN FAIL the poin...

Page 14: ...full SEG Program TYPE 3 re cycles LOOP Segment 3 now appears in menu TYPE Step STEP SEG 2 TYPE Soak SOAK Q s Ramp Q s SEG TYPE 1 s SPR PCYC 1 Soak time in minutes or continue Default 10 SINT 10 Import...

Page 15: ...ocess value SP Setpoint value hour Ramp rate adjust in hours 60 s Ramp rate adjust in minutes 1 Add new segments 1 to 126 k SPrr 100 Setpoint ramp rate Units per hour minute 0 9990 as set at Spru abov...

Page 16: ...3E SP2A and SP3A energised to mark event 2d 3E SP2A de energised SP3A energised to mark event To Return to LEVL P Press and hold t To Read Programmer memory used USEd Press Q and t together in LEVL P...

Page 17: ...ions shown as white characters on black background have Default settings EDIT COPY Q3secs PROG RUN 1 OFF PROG 1 SEG ADD SEG 5 SEG 4 SEG 3 SEG 2 KEY t OR s TO VIEW FUNCTIONS Program Segment Configurati...

Page 18: ...display reads Prog Add Press button until display reads Stu Pu Press and hold and press display reads Stu SP Press button until display reads SEG Add Press button until display reads type SPR Press a...

Page 19: ...s button display reads SEG Add Press button display reads type SPR Press and hold and press button until display reads type Step Press button display reads t SP 0 0 current set point Press and hold an...

Page 20: ...PCYC 1 Press and hold and press button until display reads PCYC Cont Press button display reads EOP 3D Press button until display reads SEG 5 Press and hold both the and buttons together for 3 seconds...

Page 21: ...or the quick programming guide Models fitted with a 24 hour or 7 day time clock require the segments to be adjusted by pulling outwards or pushing in according to requirements Each small segment repre...

Page 22: ...ed alarm option fitted to the instrument s front panel When fitted this instrument supplements the digital control on the control panel with a contact thermometer having a Red upper limit pointer and...

Page 23: ...on the cassette with the paper unrolling from the bottom With the chart retaining bracket 5 lifted feed the paper onto the sprocketed paper driving wheels 6 Close chart retaining bracket 5 and check...

Page 24: ...gainst the chart paper In this case the recorder chart motor should be operated for a few seconds until the imprint pressure is released or the imprint drive wheel 12 can be manually rotated The chart...

Page 25: ...ce engineer The exterior and interior surfaces can be maintained in as new condition using standard non abrasive or non corrosive cleaners Solvent based cleaners must not be used Switch off during all...

Page 26: ...6 1 7 If it is found subsequently that the purchaser of the equipment was responsible for the damage malfunction a charge will be made by LMS including any carriage charges 6 1 8 If it is necessary t...

Page 27: ...PARES SERVICING AND REPAIRS 6 2 Please contact LMS Ltd The Modern Forge Amherst Hill Riverhead Sevenoaks Kent TN13 2EL Tel 01732 451866 Fax 01732 450127 Email sales lms ltd uk www lms ltd uk 27 of 31...

Page 28: ...6 3 28 of 31 01 12...

Page 29: ...Fax Back Guarantee Registration To qualify for the 2nd year guarantee please fax this form duly completed to 01732 450127 Name Address Tel No Contact Name Date Purchased Model No Serial No 29 of 31 01...

Page 30: ...ent Among the services provided are Periodic service and safety inspection Calibration services to traceable standards Emergency breakdown cover Preparation for insurance inspection UKAS Accreditation...

Page 31: ...UGE DIGESTION BLOCK INCUBATOR STERILIZER CHILLER FREEZER LABORATORY FRIDGE WATER BATH CLIMATIC CHAMBERS FRIDGE FREEZER LABORATORY OVEN COOLED INCUBATOR FURNACE MEDIA PREPARATOR PLEASE LIST THESE AND A...

Reviews: