4. Route the preamplifier cable to the location of the

P/N 33217 preamplifier through conduit, if
necessary. Unscrew and remove the preamplifier
cover. Pass the cable through the PG-9 connector.

5. Using the chart below, connect the wires from the

preamplifier cable to the preamplifier.

Note: If the cable must be cut, tin the wires
before connecting.

Terminal Connection

Color

 Temp. Comp.+

Red

 Temp. Comp. -

Black

 TC GND

Green

 Preamp Power + 5V

Orange

 Preamp Power -  5V

Blue

 Temp. Comp. -

Black

 Temp. Comp. +

Red

 VpH

White

 COM

Brown

 TC GND

Green

© 1999 LMI Milton Roy - All Rights Reserved

Printed in USA

Specifications subject to change without notice.

From Temp

Comp Cable

Cable

from Controller

Fig. B

6. Connect the pH/ORP electrode BNC connector to

the mating BNC on the preamplifier .

If temperature compensation is to be used, remove
the white plastic dowel pin from the PG-7 connec
tor. Run the temperature compensation wires (red,
black and green) in through the connector and
connect to the three terminals of the electronic
module as in the chart above.