E-mail support
LMI Corporation
Phone support
(810) 714-5811
Form: CA 045
May 5, 2005
C:\Documents and Settings\timmartin\Local Settings\Temporary Internet Files\OLK9\CA
045 LMI 241-BWV2 to 440 or 501.doc
Page 12 of
12
Gap Verification
5.
Press
►►
on the keypad of the collector to change the display from G2 to G3.
6.
With the gap peg extended, observe the
value in the “Master Gages” screen of
“Gage Port G3” in the collector.
7.
Retract the gap finger; observe the
value in the “Master Gages” screen of
“Gage Port G3b” in the collector.
8.
Subtract the results of step 7 from step 6; the result must be 10.00mm +/- 0.03mm. This example
step 7 is 1.802 and step 6 is 11.802, the result is 10.000. 11.802 - 1.802 = 10.000
9.
Insert the gage into the master block
and observe the value in the “Master
Gages” screen of “Gage Port G3” in
the collector. It must be 6.00 +/-
0.03mm.
Verification Complete
Gage port “G3”