C-015 Unidentified loss
of air flow
and submit to LM fax +358 2 454 6444. After obtaining an
RMA number from LM, return the entire device with
accessories and the RMA number to LM for repair.
E-P03 No light in
handpiece
X
1. Replace the AirLED tubing with a working
polisher tubing. If this solved the problem the
polisher tubing was faulty.
C-014 Faulty AirLED
tubing.
Replace AirLED tubing (spare part 15610261).
Fill in a
indicating cause code C-017
and submit to LM fax +358 2 454 6444. After obtaining an
RMA number from LM, return the faulty tubing for
replacement.
2. If above action did not solve the problem,
request authorization to return the device to
LM-Instruments
C-007 Unidentified loss
handpiece light
Fill in
and submit to LM fax +358 2 454 6444. After obtaining an
RMA number from LM, return the entire device with
accessories and the RMA number to LM for repair.
E-P04 No/low water flow
X
1. Check that the water control on the powder
container cap is fully open. Replace the nozzle
with a working nozzle. If this solved the
problem clean the clogged nozzle.
C-013 Clogged nozzle
Clean the clogged nozzle in an ultrasonic bath or in citric acid
for 10 min. Rinse with water afterwards and dry with
compressed air.
If above did not work, submit
indicating cause code C-013 to LM fax +358
2 454 6444. After obtaining an RMA number from LM, return
the nozzle for repair.
2. Check whether any o-rings are worn
(polisher tubing connectors and bottle cap and
connectors when applicable).
C-021 O-rings
cause code C-021 and submit to LM fax +358 2 454 6444.