LLG uniSTIRRER 3 Operation Manual Download Page 4

Description     

Sets  the  temperature  parameters. 

The function “heating” is switched ON or OFF by pushing 

the temperature control knob. 

Sets the stirring speed. 

The function “Stirring” is switched ON or OFF by pushing the  

speed control knob. 

LED displays the real working state and all settings. 

When the  external temperature sensor PT 1000 is  plugged 

in, probe icon 

is lit. 

Switches the instrument ON or OFF 

 

When  heating function  was switched  ON,  the  LED displays  

the temperature setting value and shifts to real value in 5  

seconds. 

When the heating function is switched OFF and the heating 

plate temperature is still above 50 °C, LED displays “Hot”,  

otherwise LED displays OFF. 

 
 

When stirring function was switched ON, the LED displays  

the speed setting value  and flashes. The setting value does  

not flash until real speed reaches the set point. 

 
 
 
 
 
 
 

• 

Make sure the required operating voltage and power supply voltage match. 

• 

Ensure the socket is properly grounded. 

• 

Plug in the power cable, ensure the power is on and begin initialising. 

• 

Add the medium into the vessel with an appropriate stirring bar. 

• 

Place vessel on the heating plate. 

• 

Set the target stirring speed and begin. 

• 

Observe the stirring bar and LED display. 

• 

Set the target temperature and start heating. 

• 

Observe the LED display. 

• 

Stop the heating and stirring functions. 

 
If the operations above are normal the unit is ready to operate. If not, the unit may have been dam- 

aged during transit and in this case please contact your distributor. 

 

                                                   

Item 

Temperature control knob “Heat”

 
 

Speed control knob “Stir” 

 
 

LED display 

Probe 

 

Power switch 

Temperature display area (1) 

 
 
 

 
 

Speed display area (2) 

 
 
 

 
 

5.2  Operating 

Initialisation 

Summary of Contents for uniSTIRRER 3

Page 1: ...ENG LISH Operation Manual Magnetic stirrer with heating uniSTIRRER 3...

Page 2: ...result from mechanical vibrations During operation Gradually increase the speed reduce the speed if the stirring bar breaks away due to high speed or the instrument is not running smoothly or the con...

Page 3: ...t carefully and check for any damages If the instrument is faulty you should contact your distributor The packing includes 1 Magnetic Stirrer with Heating 1 Temperature sensor PT 1000 with stand 1 pow...

Page 4: ...he speed setting value and flashes The setting value does not flash until real speed reaches the set point Make sure the required operating voltage and power supply voltage match Ensure the socket is...

Page 5: ...ctual temperature of the sample in the vessel or of the surface of the heating plate The temperature differences are as following Heating plate center and outer edge The sample in the vessel and the s...

Page 6: ...f 10 rpm Troubleshooting Instrument can not be powered ON check that the power cord is plugged in Fault in power on self test switch OFF the unit then switch ON and reset the instrument to factory def...

Page 7: ...es if the instrument is opened by untrained or unqualified staff Our General terms and conditions apply Service If help is needed please contact your distributor Please provide the customer care repre...

Page 8: ...r input 5 W Motor power output 3 W Speed range 100 1500 rpm Display LED Heating plate diameter 135 mm Heating power 500 W Temperature range Room temp to 280 C in 1 C increments Safety temperature 320...

Page 9: ...Lab Logistics Group GmbH LLG GmbH Am Hambuch 1 D 53340 Meckenheim Tel 49 2225 9211 0 Fax 49 2225 9211 11 www llg de info llg de...