BUTTON ASSIGNMENTS
The maximum and minimum measured values for a measurement period can be retrieved using the max/min button
max/min button
max/min button
max/min button. When the max/min
button is actuated firstly the max symbol and the maximum measured value appear in the display. With a further actuation the minimum
value and the min symbol appear. Alternation between these two states can be carried out as often as required. If during measurement the
maximum possible measured value is exceeded (Err1), then the maximum possible measured value is displayed. If the measurement moves
out of range at the bottom end (Err2) the minimum possible measured value is displayed. If the button is pressed for approx 3 seconds, the
memory is cleared. 0.00 appears on the display as confirmation of this. If no button is actuated for approx 5 seconds the unit program reverts
into the normal measuring mode.
The hold button
hold button
hold button
hold button enables the measured value to be frozen. This is useful if measurements are being taken at locations where full attention
must be paid to the measurement site. This function is also useful for temperature comparisons over longer time periods.
The unit is switched on by actuating the on/off button
on/off button
on/off button
on/off button. When the unit is switched off the max/min store is deleted. All other settings are
retained.
BATTERY MANAGEMENT
Attention is drawn to a flat battery in 2 stages. If the battery symbol appears on the display the battery should be replaced soon. However,
measurements can still be made, and the accuracy of these is still guaranteed. If the battery voltage drops further, Err4 appears on the display.
Measurements are then no longer possible and the battery must be replaced as a matter of urgency. The battery compartment is located on
the rear face of the housing. It is opened by pressing lightly with the thumbs on the hatched surface and pulling the flap rearward.
INTERFACE
The unit uses an USB interface as standard. The software enables the display and storage of measured values and in addition certain
parameters can be adjusted. Graphical presentations of temperatures over time and imports of values into Excel
are also possible.
SENSORS
A large number of sensors are available for the DTM3000. If a particularly complicated measurement problem arises, we would be happy to
produce a sensor to your requirements at an attractive price. You can find a selection of sensors on our Internet site (www.lkmelectronic.de).
ADVICE REGARDING ACCURATE MEASUREMENTS
Thermocouples stands out for simplicity assembly, robustness and high measurement cruise.
But the accuracy is lower than the accuracy of pt100 elements.
In fact of the cold-junction compensation (Seebeck-effect) they become more mistakes too.
In the event of extreme temperature changes (e.g. moving from a warm room into a cold environment) a thermocouple hand unit should be
allowed to acclimatise for app. 10 minutes to ensure accurate measurements. For surface measurements there are special sensor designs.
The unit should be checked every 2 years. To satisfy this re
The unit should be checked every 2 years. To satisfy this re
The unit should be checked every 2 years. To satisfy this re
The unit should be checked every 2 years. To satisfy this requirement we are happy to provide you with an inexpensive works
quirement we are happy to provide you with an inexpensive works
quirement we are happy to provide you with an inexpensive works
quirement we are happy to provide you with an inexpensive works
calibration certificate.
calibration certificate.
calibration certificate.
calibration certificate.
POSSIBLE FAULTS
Symptoms
Symptoms
Symptoms
Symptoms
Possible cause
Possible cause
Possible cause
Possible cause
Error1 even with sensor
connected at room temperature
Sensor fracture
Err2 always displayed.
Short circuit in cable or connector
Display value is too low and has severe
oscillations
Moisture has penetrated into the sensor or the unit
Showing temperature change in one
direction in face of constant measuring
point temperature
Unit is not enough acclimatise,
wait app. 10 min, till the showed temperature
get in one direction constant
Subject to modifications arising from errors or technical advancements
July 2011