BUTTON ASSIGNMENTS 

 

 

The maximum and minimum measured values for a measurement period can be retrieved using the max/min button

max/min button

max/min button

max/min button.  When the max/min 

button is actuated firstly the max symbol and the maximum measured value appear in the display.  With a further actuation the minimum 
value and the min symbol appear.  Alternation between these two states can be carried out as often as required.  If during measurement the 
maximum possible measured value is exceeded (Err1), then the maximum possible measured value is displayed.  If the measurement moves 
out of range at the bottom end (Err2) the minimum possible measured value is displayed.  If the button is pressed for approx 3 seconds, the 
memory is cleared.  0.00 appears on the display as confirmation of this.  If no button is actuated for approx 5 seconds the unit program reverts 
into the normal measuring mode. 
 
The  hold  button

hold  button

hold  button

hold  button enables the measured value to be frozen.  This is useful if measurements are being taken at locations where full attention 

must be paid to the measurement site.  This function is also useful for temperature comparisons over longer time periods. 
 
The  unit  is  switched  on  by  actuating  the  on/off  button

on/off  button

on/off  button

on/off  button.  When  the  unit  is  switched  off  the  max/min  store  is  deleted.  All  other  settings  are 

retained. 
 

 

BATTERY MANAGEMENT 

 

 
Attention is drawn to a flat battery in 2 stages.  If the battery symbol appears on the display the battery should be replaced soon.  However, 
measurements can still be made, and the accuracy of these is still guaranteed.  If the battery voltage drops further, Err4 appears on the display.  
Measurements are then no longer possible and the battery must be replaced as a matter of urgency.  The battery compartment is located on 
the rear face of the housing.  It is opened by pressing lightly with the thumbs on the hatched surface and pulling the flap rearward.  

 

INTERFACE 

 

 

The unit uses an USB interface as standard.  The software enables the display and storage of measured values and in addition certain 
parameters can be adjusted.  Graphical presentations of temperatures over time and imports of values into Excel 

 are also possible. 

 

SENSORS 

 

 
 

A large number of sensors are available for the DTM3000.  If a particularly complicated measurement problem arises, we would be happy to 
produce a sensor to your requirements at an attractive price. You can find a selection of sensors on our Internet site (www.lkmelectronic.de). 

 
 

ADVICE REGARDING ACCURATE MEASUREMENTS 

 

 
 

Thermocouples stands out for simplicity assembly, robustness and high measurement cruise. 
But the accuracy is lower than the accuracy  of pt100 elements.  
In fact of the cold-junction compensation (Seebeck-effect) they become more mistakes too.  
In the event of extreme temperature changes (e.g. moving from a warm room into a cold environment) a thermocouple hand unit should be 
allowed to acclimatise for app. 10 minutes to ensure accurate measurements. For surface measurements there are special sensor designs.

 

 

The unit should be checked every 2 years. To satisfy this re

The unit should be checked every 2 years. To satisfy this re

The unit should be checked every 2 years. To satisfy this re

The unit should be checked every 2 years. To satisfy this requirement we are happy to provide you with an inexpensive works 

quirement we are happy to provide you with an inexpensive works 

quirement we are happy to provide you with an inexpensive works 

quirement we are happy to provide you with an inexpensive works 

calibration certificate.

calibration certificate.

calibration certificate.

calibration certificate.    

 

POSSIBLE FAULTS 

 

 

 

Symptoms

Symptoms

Symptoms

Symptoms    

    

Possible cause

Possible cause

Possible cause

Possible cause    

Error1 even with sensor  
connected at room temperature 

Sensor fracture 

Err2 always displayed. 

Short circuit in cable or connector 

Display value is too low and has severe 
oscillations 

Moisture has penetrated into the sensor or the unit 

Showing temperature change in one 
direction in face of constant measuring 
point temperature       

Unit is not enough acclimatise,  
wait app. 10 min, till the showed temperature                                                                                       
get in one direction constant   

 

 
 
 
 
 

Subject to modifications arising from errors or technical advancements   

 

         

 

 

           July 2011 