LK LI-IL060 User Manual Download Page 1

Summary of Contents for LI-IL060

Page 1: ......

Page 2: ...ate products In case some parts which need extra care for users we put some mark as below for the occasion Warning Mark This is mark for Dangerous Situation If users ignore this it might cause of serious personal injury or damage for products Attention Mark This is mark come up with the situation which needs extra care When users recognize this sign they have to operate more carefully 3 ...

Page 3: ...n 2 1 Naming and Function of Temperature Controller 2 2 FIX MODE One Temp Operation 2 3 Program MODE Multiple Temp Operation 2 4 Additional Function 3 Maintain 3 1 Maintaining after Use 3 2 Disorders and Solutions 3 3 After Sales Service 4 Specification 5 5 5 6 9 10 10 12 14 17 18 18 18 20 21 4 ...

Page 4: ...ontrol prevent spreading of samples inside You can observe the internal conveniently through the Glass Door Graphic LCD Display provide handy operation and control Program designed to control at multiple temperature 10 step control available PID control system with high performance microprocessor provide quick and precision temp control By Auto Tuning installed provide automatic calculation of PID...

Page 5: ...or 6 1 3 continued 4 3 9 2 6 1 7 8 5 10 11 12 13 14 1 Over Temp Limiter 2 Temperature Controller 3 Door 4 Door Handle 5 Glass Door 6 Shelf rail 7 Shelf 8 Door Packing 9 Chamber 10 Power Swtich 11 Communication Port 12 Circuit Breaker 13 Power Cord 14 Lower Cover ...

Page 6: ...tion between inside and outside of chamber prevent overheated of surface 4 Door Handle handle to open and close the door 5 Glass Door observe the internal through the Glass Door 6 Shelf rail used for fixing shelf adjustable of height of shelf 7 Shelf 2 type as wire and perforated used for put on the sample inside chamber 8 Door Packing Silicone packing resistant seal the chamber 9 Chamber space fo...

Page 7: ...ion 11 Communication Port RS485 communication port 12 Circuit Breaker Breaker of overcurrent 13 Power Cord supply electronic power to equipment 14 Lower Cover Lower part on the cover there are condenser inside 8 ...

Page 8: ...Avoid direct light Place where with less vibration and flat surface Maintain the temperature of surrounding area the product placed between 5 to 40 Maintain the humidity of surrounding area the product placed below RH 80 Avoid the place where may occur flammable gas Avoid the place where may occur noise and high frequency Avoid the place where may occur overcurrent or water leak Avoid the place wh...

Page 9: ...t on during the equipment operates 3 Heater Lamp Display quantity of output with flickering 4 Timer Lamp Light on during operate with timer 5 A T Lamp Light on during operate with auto tuning 6 UP key Use for increasing of set value and set of Fix Mode 7 Down key Use for decreasing of set value and set of Program Mode 10 1 2 3 4 5 10 6 9 7 8 11 13 17 12 15 14 16 ...

Page 10: ...ram multiple temp Mode 12 Message Displaying Window Display of the message related status of the product 13 PV Display Display current temperature of equipment 14 SV Display Display target temperature the user set 15 Timer Display Timer display remain time for operation means timer off 16 Fan Speed Display Display Step of inner circulating fan available to set 0 to 5 17 RUN STOP Display Display th...

Page 11: ...temp operation Operation Method 2 2 Check the mode status and if not set as FIX MODE press UP KEY for 3 seconds to change FIX MODE SV Set Mode Set SV Value Timer Set Mode Set Time Value Operation Method Continued Fix Mode Stand By ...

Page 12: ...2 Operation Start Operation 13 Fan Speed Set Mode Set Speed Value Fix Mode Stand By 2sec Lower setting of fan speed can cause for bigger temperature difference inside chamber ...

Page 13: ...de status and if not set as PROGRAM MODE press DOWN KEY for 3 seconds to change PROGRAM MODE End Step Set mode Set number of steps As number of steps set user can set temperature steps Step 1 SV Set mode Set SV Value for Step 1 Program Mode Stand By Program Mode Setting Method Continued ...

Page 14: ...p 1 Timer Set mode Set time value for Step 1 Step 1 Fan Set Mode Set speed value for Step 1 Step 2 SV Set mode Set SV Value for Step 2 Step 2 Timer Set mode Set time value for Step 2 Program Mode Setting Method Continued ...

Page 15: ...2 Operation 16 Step 1 Fan Set Mode Set speed value for Step 2 Program Mode Stand By With same method set values till the last step in order of temp time fan speed 2sec Start Operation ...

Page 16: ...if user wish to same experiment once again the user need only tuning once Auto Tuning is only available during RUN session Fix control mode Start with Auto Tuning message After the Auto Tuning the equipment will operate with set temperature During Auto Tuning due to calculate in order to reach set temperature Heater gives 100 output so temperature go higher than set value In case stop Auto Tuning ...

Page 17: ...e compressor Condenser once in two months Do not use strong acid alkaline or volatility solution for cleaning the equipment Also completely dry the equipment after cleaning Disorder and Solution 3 2 3 2 1 Power On Disorder 1 Please check power supply 2 Please check if Power Switch is ON 3 Please check if Power Cord is well connected with main body 4 Please check if the consent s breaker is ON 5 If...

Page 18: ... solution above contact our Technical Support Department 3 2 4 Error Messege 1 Door Open When door remains open 2 Sensor Open error Occur when there s some problem with sensor contact our Technical Support Department 3 LBA Control Routine Breaking Down Alarm Please set the Over Temp Limiter set 10 higher than treat value If the error message still comes up after set contact our Technical Support D...

Page 19: ...illed usage cannot be get warranty service 3 Applying A S Firstly contact to our Technical Support Department or Supplier of the products and inform detailed sympathy with contact of user by mail or fax After receipt of A S inquiry our technician quotes and user decides after get quoted The product after 2 weeks from receipt of A S inquiry without response will be return to the user Technical Supp...

Page 20: ...ontrol Autotuning GLCD Graphic LCD Resolution 0 01 Temperature Range 0 80 Accuracy at 37 0 05 Uniformity at 37 0 54 0 63 0 76 0 68 Dimension Interior w d h 420 350 450 mm 505 505 610 mm 505 505 1000 mm 720 620 1000 mm Exterior w d h 530 590 1120 mm 640 830 1310 mm 640 830 1700 mm 850 950 1750 mm Material Interior Stainless Steel SUS304 External Powder Coated Steel Electric Supply Power 1 Phase 220...

Page 21: ... LKLABKorea ...

Page 22: ......

Reviews: